- Delete placeholder files (example.py, api_reference.md, example_asset.txt)
- Create 3 real reference files with MDR compliance content:
- mdr-classification-guide.md: Annex VIII rules, MDCG 2019-11, examples
- clinical-evidence-requirements.md: CER structure, PMCF, literature review
- technical-documentation-templates.md: Annex II/III, GSPR, DHF templates
- Create mdr_gap_analyzer.py (~350 lines): Interactive gap analysis tool
- Rewrite SKILL.md with trigger phrases, TOC, numbered workflows
- Add 3 concrete classification examples with rationales
- Add validation checkpoints to all workflows
- Remove marketing language ("Expert-level...comprehensive knowledge")
Co-authored-by: Claude Opus 4.5 <noreply@anthropic.com>
This commit is contained in:
@@ -1,195 +1,335 @@
|
||||
---
|
||||
name: mdr-745-specialist
|
||||
description: EU MDR 2017/745 regulation specialist and consultant for medical device requirement management. Provides comprehensive MDR compliance expertise, gap analysis, technical documentation guidance, clinical evidence requirements, and post-market surveillance implementation. Use for MDR compliance assessment, classification decisions, technical file preparation, and regulatory requirement interpretation.
|
||||
description: EU MDR 2017/745 compliance specialist for medical device classification, technical documentation, clinical evidence, and post-market surveillance. Covers Annex VIII classification rules, Annex II/III technical files, Annex XIV clinical evaluation, and EUDAMED integration.
|
||||
triggers:
|
||||
- MDR compliance
|
||||
- EU MDR
|
||||
- medical device classification
|
||||
- Annex VIII
|
||||
- technical documentation
|
||||
- clinical evaluation
|
||||
- PMCF
|
||||
- EUDAMED
|
||||
- UDI
|
||||
- notified body
|
||||
---
|
||||
|
||||
# Senior MDR 2017/745 Specialist and Consultant
|
||||
# MDR 2017/745 Specialist
|
||||
|
||||
Expert-level EU MDR 2017/745 compliance specialist with comprehensive knowledge of medical device regulation requirements, technical documentation, clinical evidence, and post-market surveillance obligations.
|
||||
EU MDR compliance patterns for medical device classification, technical documentation, and clinical evidence.
|
||||
|
||||
## Core MDR Competencies
|
||||
---
|
||||
|
||||
### 1. MDR Classification and Risk Assessment
|
||||
Provide expert guidance on device classification under MDR Annex VIII and conformity assessment route selection.
|
||||
## Table of Contents
|
||||
|
||||
**Classification Decision Framework:**
|
||||
1. **Preliminary Classification Assessment**
|
||||
- Apply MDR Annex VIII classification rules
|
||||
- Consider device duration, invasiveness, and body system interaction
|
||||
- Evaluate software classification per MDCG 2019-11
|
||||
- **Decision Point**: Determine appropriate classification class (I, IIa, IIb, III)
|
||||
- [Device Classification Workflow](#device-classification-workflow)
|
||||
- [Technical Documentation](#technical-documentation)
|
||||
- [Clinical Evidence](#clinical-evidence)
|
||||
- [Post-Market Surveillance](#post-market-surveillance)
|
||||
- [EUDAMED and UDI](#eudamed-and-udi)
|
||||
- [Reference Documentation](#reference-documentation)
|
||||
- [Tools](#tools)
|
||||
|
||||
2. **Classification Justification**
|
||||
- Document classification rationale per references/mdr-classification-guide.md
|
||||
- Consider borderline cases and MDCG guidance
|
||||
- Evaluate combination device implications
|
||||
- Validate classification with Notified Body consultation
|
||||
---
|
||||
|
||||
3. **Conformity Assessment Route Selection**
|
||||
- **Class I**: Self-certification under Annex II
|
||||
- **Class IIa**: Module C2 + Annex V (Notified Body involvement)
|
||||
- **Class IIb**: Module B + C or D (Type examination + production)
|
||||
- **Class III**: Module B + C or D (Full quality assurance)
|
||||
## Device Classification Workflow
|
||||
|
||||
### 2. Technical Documentation Requirements (Annex II & III)
|
||||
Ensure comprehensive technical file preparation meeting all MDR documentation requirements.
|
||||
Classify device under MDR Annex VIII:
|
||||
|
||||
1. Identify device duration (transient, short-term, long-term)
|
||||
2. Determine invasiveness level (non-invasive, body orifice, surgical)
|
||||
3. Assess body system contact (CNS, cardiac, other)
|
||||
4. Check if active device (energy dependent)
|
||||
5. Apply classification rules 1-22
|
||||
6. For software, apply MDCG 2019-11 algorithm
|
||||
7. Document classification rationale
|
||||
8. **Validation:** Classification confirmed with Notified Body
|
||||
|
||||
### Classification Matrix
|
||||
|
||||
| Factor | Class I | Class IIa | Class IIb | Class III |
|
||||
|--------|---------|-----------|-----------|-----------|
|
||||
| Duration | Any | Short-term | Long-term | Long-term |
|
||||
| Invasiveness | Non-invasive | Body orifice | Surgical | Implantable |
|
||||
| System | Any | Non-critical | Critical organs | CNS/cardiac |
|
||||
| Risk | Lowest | Low-medium | Medium-high | Highest |
|
||||
|
||||
### Software Classification (MDCG 2019-11)
|
||||
|
||||
| Information Use | Condition Severity | Class |
|
||||
|-----------------|-------------------|-------|
|
||||
| Informs decision | Non-serious | IIa |
|
||||
| Informs decision | Serious | IIb |
|
||||
| Drives/treats | Critical | III |
|
||||
|
||||
### Classification Examples
|
||||
|
||||
**Example 1: Absorbable Surgical Suture**
|
||||
- Rule 8 (implantable, long-term)
|
||||
- Duration: > 30 days (absorbed)
|
||||
- Contact: General tissue
|
||||
- Classification: **Class IIb**
|
||||
|
||||
**Example 2: AI Diagnostic Software**
|
||||
- Rule 11 + MDCG 2019-11
|
||||
- Function: Diagnoses serious condition
|
||||
- Classification: **Class IIb**
|
||||
|
||||
**Example 3: Cardiac Pacemaker**
|
||||
- Rule 8 (implantable)
|
||||
- Contact: Central circulatory system
|
||||
- Classification: **Class III**
|
||||
|
||||
---
|
||||
|
||||
## Technical Documentation
|
||||
|
||||
Prepare technical file per Annex II and III:
|
||||
|
||||
1. Create device description (variants, accessories, intended purpose)
|
||||
2. Develop labeling (Article 13 requirements, IFU)
|
||||
3. Document design and manufacturing process
|
||||
4. Complete GSPR compliance matrix
|
||||
5. Prepare benefit-risk analysis
|
||||
6. Compile verification and validation evidence
|
||||
7. Integrate risk management file (ISO 14971)
|
||||
8. **Validation:** Technical file reviewed for completeness
|
||||
|
||||
### Technical File Structure
|
||||
|
||||
**Technical Documentation Structure:**
|
||||
```
|
||||
ANNEX II TECHNICAL DOCUMENTATION
|
||||
├── General Information
|
||||
│ ├── Device identification and UDI-DI
|
||||
│ ├── Manufacturer and authorized representative info
|
||||
│ ├── Intended purpose and clinical condition
|
||||
│ └── Device description and variants
|
||||
├── Information to be Supplied by Manufacturer
|
||||
│ ├── Label and instructions for use
|
||||
│ ├── Clinical evaluation and post-market clinical follow-up
|
||||
│ ├── Risk management documentation
|
||||
│ └── Product verification and validation
|
||||
├── Design and Manufacturing Information
|
||||
│ ├── Quality management system documentation
|
||||
│ ├── Design and development process
|
||||
│ ├── Manufacturing process description
|
||||
│ └── Identification and traceability procedures
|
||||
└── General Safety and Performance Requirements
|
||||
├── Solutions adopted for GSPR compliance
|
||||
├── Benefit-risk analysis and risk management
|
||||
├── Product lifecycle and post-market surveillance
|
||||
└── Clinical evidence and evaluation
|
||||
├── Device description and UDI-DI
|
||||
├── Label and instructions for use
|
||||
├── Design and manufacturing info
|
||||
├── GSPR compliance matrix
|
||||
├── Benefit-risk analysis
|
||||
├── Verification and validation
|
||||
└── Clinical evaluation report
|
||||
```
|
||||
|
||||
### 3. Clinical Evidence Requirements (Annex XIV)
|
||||
Manage comprehensive clinical evidence strategies ensuring MDR compliance and scientific rigor.
|
||||
### GSPR Compliance Checklist
|
||||
|
||||
**Clinical Evidence Pathway Selection:**
|
||||
1. **Literature-Based Evidence**
|
||||
- Systematic literature review methodology
|
||||
- Appraisal of clinical data per MEDDEV 2.7/1 rev.4
|
||||
- Gap analysis and additional evidence requirements
|
||||
- **Decision Point**: Determine if literature is sufficient or clinical investigation required
|
||||
| Requirement | Evidence | Status |
|
||||
|-------------|----------|--------|
|
||||
| Safe design (GSPR 1-3) | Risk management file | ☐ |
|
||||
| Chemical properties (GSPR 10.1) | Biocompatibility report | ☐ |
|
||||
| Infection risk (GSPR 10.2) | Sterilization validation | ☐ |
|
||||
| Software requirements (GSPR 17) | IEC 62304 documentation | ☐ |
|
||||
| Labeling (GSPR 23) | Label artwork, IFU | ☐ |
|
||||
|
||||
2. **Clinical Investigation Requirements**
|
||||
- **For significant changes** or **novel devices**
|
||||
- **For Class III implantable devices** (Article 61)
|
||||
- Clinical investigation plan development
|
||||
- Ethics committee and competent authority approvals
|
||||
### Conformity Assessment Routes
|
||||
|
||||
3. **Post-Market Clinical Follow-up (PMCF)**
|
||||
- **PMCF Plan** development per Annex XIV Part B
|
||||
- **PMCF Evaluation Report** (PMCF-ER) preparation
|
||||
- Clinical evaluation report updating requirements
|
||||
- Integration with post-market surveillance system
|
||||
| Class | Route | NB Involvement |
|
||||
|-------|-------|----------------|
|
||||
| I | Annex II self-declaration | None |
|
||||
| Is/Im | Annex II + IX/XI | Sterile/measuring aspects |
|
||||
| IIa | Annex II + IX or XI | Product or QMS |
|
||||
| IIb | Annex IX + X or X + XI | Type exam + production |
|
||||
| III | Annex IX + X | Full QMS + type exam |
|
||||
|
||||
### 4. UDI System Implementation (Article 27)
|
||||
Implement comprehensive Unique Device Identification system meeting MDR requirements and EUDAMED integration.
|
||||
---
|
||||
|
||||
**UDI Implementation Workflow:**
|
||||
1. **UDI Strategy Development**
|
||||
- UDI-DI assignment for device variants
|
||||
- UDI-PI requirements for higher risk devices
|
||||
- EUDAMED registration timeline planning
|
||||
- Labeling compliance verification
|
||||
## Clinical Evidence
|
||||
|
||||
2. **EUDAMED Registration**
|
||||
- **Actor registration** (manufacturers, authorized representatives)
|
||||
- **Device registration** and UDI-DI assignment
|
||||
- **Certificate registration** (Notified Body certificates)
|
||||
- **Clinical investigation** and serious incident reporting
|
||||
Develop clinical evidence strategy per Annex XIV:
|
||||
|
||||
## MDR Compliance Management
|
||||
1. Define clinical claims and endpoints
|
||||
2. Conduct systematic literature search
|
||||
3. Appraise clinical data quality
|
||||
4. Assess equivalence (technical, biological, clinical)
|
||||
5. Identify evidence gaps
|
||||
6. Determine if clinical investigation required
|
||||
7. Prepare Clinical Evaluation Report (CER)
|
||||
8. **Validation:** CER reviewed by qualified evaluator
|
||||
|
||||
### Gap Analysis and Transition Planning
|
||||
Conduct systematic gap assessments against current MDR requirements and develop comprehensive transition strategies.
|
||||
### Evidence Requirements by Class
|
||||
|
||||
**Gap Analysis Framework:**
|
||||
1. **Current State Assessment**
|
||||
- Existing QMS compliance evaluation
|
||||
- Technical documentation gap identification
|
||||
- Clinical evidence adequacy assessment
|
||||
- Post-market surveillance system review
|
||||
| Class | Minimum Evidence | Investigation |
|
||||
|-------|------------------|---------------|
|
||||
| I | Risk-benefit analysis | Not typically required |
|
||||
| IIa | Literature + post-market | May be required |
|
||||
| IIb | Systematic literature review | Often required |
|
||||
| III | Comprehensive clinical data | Required (Article 61) |
|
||||
|
||||
2. **MDR Requirement Mapping**
|
||||
- **For existing devices**: Legacy directive vs. MDR requirements
|
||||
- **For new devices**: Full MDR compliance roadmap
|
||||
- **For software**: Software-specific MDR requirements per MDCG guidance
|
||||
- Resource and timeline impact assessment
|
||||
### Clinical Evaluation Report Structure
|
||||
|
||||
### Post-Market Surveillance (Chapter VII)
|
||||
Establish robust post-market surveillance systems meeting MDR requirements for continuous safety monitoring.
|
||||
```
|
||||
CER CONTENTS
|
||||
├── Executive summary
|
||||
├── Device scope and intended purpose
|
||||
├── Clinical background (state of the art)
|
||||
├── Literature search methodology
|
||||
├── Data appraisal and analysis
|
||||
├── Safety and performance conclusions
|
||||
├── Benefit-risk determination
|
||||
└── PMCF plan summary
|
||||
```
|
||||
|
||||
**PMS System Components:**
|
||||
- **PMS Plan** development per Article 84
|
||||
- **Periodic Safety Update Report (PSUR)** preparation
|
||||
- **Serious incident reporting** to competent authorities
|
||||
- **Field safety corrective actions (FSCA)** management
|
||||
- **Trend reporting** and signal detection
|
||||
### Qualified Evaluator Requirements
|
||||
|
||||
### Economic Operator Obligations
|
||||
Ensure compliance with expanded economic operator responsibilities under MDR.
|
||||
- Medical degree or equivalent healthcare qualification
|
||||
- 4+ years clinical experience in relevant field
|
||||
- Training in clinical evaluation methodology
|
||||
- Understanding of MDR requirements
|
||||
|
||||
**Key Obligations Management:**
|
||||
- **Manufacturer obligations** (Article 10)
|
||||
- **Authorized representative duties** (Article 11)
|
||||
- **Importer responsibilities** (Article 13)
|
||||
- **Distributor obligations** (Article 14)
|
||||
- **Person responsible for regulatory compliance** (Article 15)
|
||||
---
|
||||
|
||||
## Post-Market Surveillance
|
||||
|
||||
Establish PMS system per Chapter VII:
|
||||
|
||||
1. Develop PMS plan (Article 84)
|
||||
2. Define data collection methods
|
||||
3. Establish complaint handling procedures
|
||||
4. Create vigilance reporting process
|
||||
5. Plan Periodic Safety Update Reports (PSUR)
|
||||
6. Integrate with PMCF activities
|
||||
7. Define trend analysis and signal detection
|
||||
8. **Validation:** PMS system audited annually
|
||||
|
||||
### PMS System Components
|
||||
|
||||
| Component | Requirement | Frequency |
|
||||
|-----------|-------------|-----------|
|
||||
| PMS Plan | Article 84 | Maintain current |
|
||||
| PSUR | Class IIa and higher | Per class schedule |
|
||||
| PMCF Plan | Annex XIV Part B | Update with CER |
|
||||
| PMCF Report | Annex XIV Part B | Annual (Class III) |
|
||||
| Vigilance | Articles 87-92 | As events occur |
|
||||
|
||||
### PSUR Schedule
|
||||
|
||||
| Class | Frequency |
|
||||
|-------|-----------|
|
||||
| Class III | Annual |
|
||||
| Class IIb implantable | Annual |
|
||||
| Class IIb | Every 2 years |
|
||||
| Class IIa | When necessary |
|
||||
|
||||
### Serious Incident Reporting
|
||||
|
||||
| Timeline | Requirement |
|
||||
|----------|-------------|
|
||||
| 2 days | Serious public health threat |
|
||||
| 10 days | Death or serious deterioration |
|
||||
| 15 days | Other serious incidents |
|
||||
|
||||
---
|
||||
|
||||
## EUDAMED and UDI
|
||||
|
||||
Implement UDI system per Article 27:
|
||||
|
||||
1. Obtain issuing entity code (GS1, HIBCC, ICCBBA)
|
||||
2. Assign UDI-DI to each device variant
|
||||
3. Assign UDI-PI (production identifier)
|
||||
4. Apply UDI carrier to labels (AIDC + HRI)
|
||||
5. Register actor in EUDAMED
|
||||
6. Register devices in EUDAMED
|
||||
7. Upload certificates when available
|
||||
8. **Validation:** UDI verified on sample labels
|
||||
|
||||
### EUDAMED Modules
|
||||
|
||||
| Module | Content | Actor |
|
||||
|--------|---------|-------|
|
||||
| Actor | Company registration | Manufacturer, AR |
|
||||
| UDI/Device | Device and variant data | Manufacturer |
|
||||
| Certificates | NB certificates | Notified Body |
|
||||
| Clinical Investigation | Study registration | Sponsor |
|
||||
| Vigilance | Incident reports | Manufacturer |
|
||||
| Market Surveillance | Authority actions | Competent Authority |
|
||||
|
||||
### UDI Label Requirements
|
||||
|
||||
Required elements per Article 13:
|
||||
|
||||
- [ ] UDI-DI (device identifier)
|
||||
- [ ] UDI-PI (production identifier) for Class II+
|
||||
- [ ] AIDC format (barcode/RFID)
|
||||
- [ ] HRI format (human-readable)
|
||||
- [ ] Manufacturer name and address
|
||||
- [ ] Lot/serial number
|
||||
- [ ] Expiration date (if applicable)
|
||||
|
||||
---
|
||||
|
||||
## Reference Documentation
|
||||
|
||||
### MDR Classification Guide
|
||||
|
||||
`references/mdr-classification-guide.md` contains:
|
||||
|
||||
- Complete Annex VIII classification rules (Rules 1-22)
|
||||
- Software classification per MDCG 2019-11
|
||||
- Worked classification examples
|
||||
- Conformity assessment route selection
|
||||
|
||||
### Clinical Evidence Requirements
|
||||
|
||||
`references/clinical-evidence-requirements.md` contains:
|
||||
|
||||
- Clinical evidence framework and hierarchy
|
||||
- Literature search methodology
|
||||
- Clinical Evaluation Report structure
|
||||
- PMCF plan and evaluation report guidance
|
||||
|
||||
### Technical Documentation Templates
|
||||
|
||||
`references/technical-documentation-templates.md` contains:
|
||||
|
||||
- Annex II and III content requirements
|
||||
- Design History File structure
|
||||
- GSPR compliance matrix template
|
||||
- Declaration of Conformity template
|
||||
- Notified Body submission checklist
|
||||
|
||||
---
|
||||
|
||||
## Tools
|
||||
|
||||
### MDR Gap Analyzer
|
||||
|
||||
```bash
|
||||
# Quick gap analysis
|
||||
python scripts/mdr_gap_analyzer.py --device "Device Name" --class IIa
|
||||
|
||||
# JSON output for integration
|
||||
python scripts/mdr_gap_analyzer.py --device "Device Name" --class III --output json
|
||||
|
||||
# Interactive assessment
|
||||
python scripts/mdr_gap_analyzer.py --interactive
|
||||
```
|
||||
|
||||
Analyzes device against MDR requirements, identifies compliance gaps, generates prioritized recommendations.
|
||||
|
||||
**Output includes:**
|
||||
- Requirements checklist by category
|
||||
- Gap identification with priorities
|
||||
- Critical gap highlighting
|
||||
- Compliance roadmap recommendations
|
||||
|
||||
---
|
||||
|
||||
## Notified Body Interface
|
||||
|
||||
### Notified Body Selection and Management
|
||||
Provide strategic guidance on Notified Body selection and relationship management throughout the conformity assessment process.
|
||||
### Selection Criteria
|
||||
|
||||
**Notified Body Engagement Strategy:**
|
||||
1. **Selection Criteria Assessment**
|
||||
- Technical competency evaluation
|
||||
- Capacity and timeline considerations
|
||||
- Geographic scope and market access
|
||||
- Fee structure and commercial terms
|
||||
| Factor | Considerations |
|
||||
|--------|----------------|
|
||||
| Designation scope | Covers your device type |
|
||||
| Capacity | Timeline for initial audit |
|
||||
| Geographic reach | Markets you need to access |
|
||||
| Technical expertise | Experience with your technology |
|
||||
| Fee structure | Transparency, predictability |
|
||||
|
||||
2. **Pre-submission Activities**
|
||||
- Pre-submission meetings and consultations
|
||||
- Technical documentation readiness assessment
|
||||
- Timeline and milestone planning
|
||||
- **Decision Point**: Determine submission readiness and timing
|
||||
### Pre-Submission Checklist
|
||||
|
||||
### Audit and Assessment Management
|
||||
Coordinate Notified Body audits and assessments ensuring successful outcomes and certificate maintenance.
|
||||
|
||||
**Audit Preparation Protocol:**
|
||||
- **Documentation preparation** and organization
|
||||
- **Personnel training** and role assignment
|
||||
- **Facility readiness** and compliance verification
|
||||
- **Mock audit** execution and improvement implementation
|
||||
|
||||
## Regulatory Intelligence and Updates
|
||||
|
||||
### MDR Guidance Monitoring
|
||||
Maintain current awareness of evolving MDR guidance and regulatory expectations.
|
||||
|
||||
**Guidance Tracking System:**
|
||||
- **MDCG guidance** monitoring and impact assessment
|
||||
- **Notified Body guidance** evaluation and implementation
|
||||
- **Competent authority positions** and national implementations
|
||||
- **Industry best practices** and lessons learned integration
|
||||
|
||||
## Resources
|
||||
|
||||
### scripts/
|
||||
- `mdr-gap-analysis.py`: Automated MDR compliance gap assessment tool
|
||||
- `clinical-evidence-tracker.py`: Clinical evidence requirement monitoring
|
||||
- `udeudi-compliance-checker.py`: UDI and EUDAMED compliance verification
|
||||
- `pms-reporting-automation.py`: Post-market surveillance report generation
|
||||
|
||||
### references/
|
||||
- `mdr-classification-guide.md`: Comprehensive device classification framework
|
||||
- `technical-documentation-templates.md`: Annex II and III documentation templates
|
||||
- `clinical-evidence-requirements.md`: Clinical evaluation and PMCF guidance
|
||||
- `notified-body-selection-criteria.md`: NB evaluation and selection framework
|
||||
- `mdcg-guidance-library.md`: Current MDCG guidance compilation
|
||||
|
||||
### assets/
|
||||
- `mdr-templates/`: Technical file, clinical evaluation, and PMS plan templates
|
||||
- `gap-analysis-checklists/`: MDR compliance assessment tools
|
||||
- `eudamed-forms/`: EUDAMED registration and reporting templates
|
||||
- `training-materials/`: MDR training presentations and compliance guides
|
||||
- [ ] Technical documentation complete
|
||||
- [ ] GSPR matrix fully addressed
|
||||
- [ ] Risk management file current
|
||||
- [ ] Clinical evaluation report complete
|
||||
- [ ] QMS (ISO 13485) certified
|
||||
- [ ] Labeling and IFU finalized
|
||||
- [ ] **Validation:** Internal gap assessment complete
|
||||
|
||||
@@ -1,24 +0,0 @@
|
||||
# Example Asset File
|
||||
|
||||
This placeholder represents where asset files would be stored.
|
||||
Replace with actual asset files (templates, images, fonts, etc.) or delete if not needed.
|
||||
|
||||
Asset files are NOT intended to be loaded into context, but rather used within
|
||||
the output Claude produces.
|
||||
|
||||
Example asset files from other skills:
|
||||
- Brand guidelines: logo.png, slides_template.pptx
|
||||
- Frontend builder: hello-world/ directory with HTML/React boilerplate
|
||||
- Typography: custom-font.ttf, font-family.woff2
|
||||
- Data: sample_data.csv, test_dataset.json
|
||||
|
||||
## Common Asset Types
|
||||
|
||||
- Templates: .pptx, .docx, boilerplate directories
|
||||
- Images: .png, .jpg, .svg, .gif
|
||||
- Fonts: .ttf, .otf, .woff, .woff2
|
||||
- Boilerplate code: Project directories, starter files
|
||||
- Icons: .ico, .svg
|
||||
- Data files: .csv, .json, .xml, .yaml
|
||||
|
||||
Note: This is a text placeholder. Actual assets can be any file type.
|
||||
@@ -1,34 +0,0 @@
|
||||
# Reference Documentation for Mdr 745 Specialist
|
||||
|
||||
This is a placeholder for detailed reference documentation.
|
||||
Replace with actual reference content or delete if not needed.
|
||||
|
||||
Example real reference docs from other skills:
|
||||
- product-management/references/communication.md - Comprehensive guide for status updates
|
||||
- product-management/references/context_building.md - Deep-dive on gathering context
|
||||
- bigquery/references/ - API references and query examples
|
||||
|
||||
## When Reference Docs Are Useful
|
||||
|
||||
Reference docs are ideal for:
|
||||
- Comprehensive API documentation
|
||||
- Detailed workflow guides
|
||||
- Complex multi-step processes
|
||||
- Information too lengthy for main SKILL.md
|
||||
- Content that's only needed for specific use cases
|
||||
|
||||
## Structure Suggestions
|
||||
|
||||
### API Reference Example
|
||||
- Overview
|
||||
- Authentication
|
||||
- Endpoints with examples
|
||||
- Error codes
|
||||
- Rate limits
|
||||
|
||||
### Workflow Guide Example
|
||||
- Prerequisites
|
||||
- Step-by-step instructions
|
||||
- Common patterns
|
||||
- Troubleshooting
|
||||
- Best practices
|
||||
@@ -0,0 +1,303 @@
|
||||
# Clinical Evidence Requirements
|
||||
|
||||
MDR Annex XIV clinical evaluation and post-market clinical follow-up guidance.
|
||||
|
||||
---
|
||||
|
||||
## Table of Contents
|
||||
|
||||
- [Clinical Evidence Framework](#clinical-evidence-framework)
|
||||
- [Clinical Evaluation Process](#clinical-evaluation-process)
|
||||
- [Literature-Based Evidence](#literature-based-evidence)
|
||||
- [Clinical Investigation Requirements](#clinical-investigation-requirements)
|
||||
- [Post-Market Clinical Follow-up](#post-market-clinical-follow-up)
|
||||
|
||||
---
|
||||
|
||||
## Clinical Evidence Framework
|
||||
|
||||
### Evidence Hierarchy
|
||||
|
||||
| Evidence Type | Strength | When to Use |
|
||||
|---------------|----------|-------------|
|
||||
| Randomized Controlled Trial | Highest | Novel Class III, high-risk claims |
|
||||
| Prospective cohort study | High | New technology, performance claims |
|
||||
| Retrospective analysis | Medium | Established technology, equivalence |
|
||||
| Literature review | Medium | Well-characterized, equivalent devices |
|
||||
| Expert opinion | Low | Supportive only, not primary |
|
||||
|
||||
### Evidence Requirements by Class
|
||||
|
||||
| Class | Minimum Evidence | Clinical Investigation |
|
||||
|-------|------------------|------------------------|
|
||||
| I | Risk-benefit analysis | Not typically required |
|
||||
| IIa | Literature + post-market data | May be required for novel tech |
|
||||
| IIb | Systematic literature review | Often required for claims |
|
||||
| III | Comprehensive clinical data | Required unless equivalent |
|
||||
|
||||
### Clinical Evidence Pathway
|
||||
|
||||
Determine evidence strategy:
|
||||
|
||||
1. Assess device classification and risk level
|
||||
2. Evaluate claim significance (diagnostic, therapeutic)
|
||||
3. Determine if equivalence can be demonstrated
|
||||
4. Identify available literature and clinical data
|
||||
5. Assess gaps requiring additional investigation
|
||||
6. Develop PMCF plan for ongoing evidence
|
||||
7. **Validation:** Evidence strategy approved by Notified Body
|
||||
|
||||
---
|
||||
|
||||
## Clinical Evaluation Process
|
||||
|
||||
### Clinical Evaluation Workflow
|
||||
|
||||
Execute clinical evaluation per Annex XIV Part A:
|
||||
|
||||
1. Identify relevant safety and performance data
|
||||
2. Define scope and search strategy
|
||||
3. Conduct systematic literature search
|
||||
4. Appraise and analyze clinical data
|
||||
5. Assess benefit-risk profile
|
||||
6. Document conclusions in Clinical Evaluation Report (CER)
|
||||
7. Plan post-market clinical follow-up
|
||||
8. **Validation:** CER reviewed by qualified clinical evaluator
|
||||
|
||||
### Clinical Evaluation Report Structure
|
||||
|
||||
```
|
||||
CLINICAL EVALUATION REPORT (CER)
|
||||
├── 1. Executive Summary
|
||||
│ ├── Device description and intended purpose
|
||||
│ ├── Conclusions on safety and performance
|
||||
│ └── Benefit-risk conclusion
|
||||
├── 2. Scope of Clinical Evaluation
|
||||
│ ├── Device identification
|
||||
│ ├── Clinical claims to be evaluated
|
||||
│ └── Equivalence assessment (if applicable)
|
||||
├── 3. Clinical Background
|
||||
│ ├── Disease/condition overview
|
||||
│ ├── Current treatment options
|
||||
│ └── State of the art
|
||||
├── 4. Clinical Data Sources
|
||||
│ ├── Pre-clinical data (bench, animal)
|
||||
│ ├── Clinical investigation data
|
||||
│ ├── Literature search methodology
|
||||
│ └── Post-market surveillance data
|
||||
├── 5. Data Appraisal
|
||||
│ ├── Study quality assessment
|
||||
│ ├── Relevance to subject device
|
||||
│ └── Data contribution to evaluation
|
||||
├── 6. Data Analysis
|
||||
│ ├── Safety analysis
|
||||
│ ├── Performance analysis
|
||||
│ └── Benefit-risk determination
|
||||
├── 7. Conclusions
|
||||
│ ├── Clinical evidence summary
|
||||
│ ├── Residual risks
|
||||
│ └── PMCF requirements
|
||||
└── 8. PMCF Plan Summary
|
||||
├── Data gaps identified
|
||||
├── PMCF activities planned
|
||||
└── Update schedule
|
||||
```
|
||||
|
||||
### Qualified Clinical Evaluator
|
||||
|
||||
Requirements per Annex XIV:
|
||||
|
||||
- Medical degree or equivalent healthcare qualification
|
||||
- 4+ years clinical experience in relevant field OR
|
||||
- Research background in relevant domain
|
||||
- Training in clinical evaluation methodology
|
||||
- Understanding of MDR requirements
|
||||
|
||||
---
|
||||
|
||||
## Literature-Based Evidence
|
||||
|
||||
### Literature Search Strategy
|
||||
|
||||
Execute systematic literature review:
|
||||
|
||||
1. Define PICO question (Population, Intervention, Comparison, Outcome)
|
||||
2. Develop search string with Boolean operators
|
||||
3. Select databases (PubMed, Embase, Cochrane, etc.)
|
||||
4. Set date range and language filters
|
||||
5. Execute search and document results
|
||||
6. Screen abstracts and full texts
|
||||
7. **Validation:** Reproducible search, documented exclusion criteria
|
||||
|
||||
### Database Selection
|
||||
|
||||
| Database | Coverage | Best For |
|
||||
|----------|----------|----------|
|
||||
| PubMed/MEDLINE | Biomedical literature | Primary clinical data |
|
||||
| Embase | Drugs, devices, biomedical | European studies |
|
||||
| Cochrane Library | Systematic reviews | Meta-analyses |
|
||||
| CINAHL | Nursing, allied health | User studies |
|
||||
| IEEE Xplore | Engineering | Technical performance |
|
||||
| Manufacturer data | Proprietary | Direct device data |
|
||||
|
||||
### Data Appraisal Criteria
|
||||
|
||||
Evaluate each source:
|
||||
|
||||
| Criterion | Assessment | Score |
|
||||
|-----------|------------|-------|
|
||||
| Study design | RCT > cohort > case series | 1-5 |
|
||||
| Sample size | Statistical power adequate | 1-5 |
|
||||
| Follow-up duration | Sufficient for outcomes | 1-5 |
|
||||
| Population relevance | Matches IFU population | 1-5 |
|
||||
| Device equivalence | Technical, biological, clinical | 1-5 |
|
||||
| Bias risk | Low/medium/high | 1-5 |
|
||||
|
||||
### Equivalence Assessment
|
||||
|
||||
Demonstrate equivalence per MDCG 2020-5:
|
||||
|
||||
**Technical equivalence:**
|
||||
- Similar design
|
||||
- Same materials
|
||||
- Same specifications
|
||||
- Same manufacturing process
|
||||
|
||||
**Biological equivalence:**
|
||||
- Same tissue contact
|
||||
- Same biocompatibility
|
||||
- Same sterilization method
|
||||
|
||||
**Clinical equivalence:**
|
||||
- Same intended purpose
|
||||
- Same clinical condition
|
||||
- Same patient population
|
||||
- Same user (professional/lay)
|
||||
|
||||
---
|
||||
|
||||
## Clinical Investigation Requirements
|
||||
|
||||
### When Investigation is Required
|
||||
|
||||
Clinical investigation mandatory:
|
||||
|
||||
- [ ] Class III implantable devices (Article 61(4))
|
||||
- [ ] Novel technology without equivalent
|
||||
- [ ] Significant modification to existing device
|
||||
- [ ] New clinical claims not supported by literature
|
||||
- [ ] Addressing gaps identified in clinical evaluation
|
||||
|
||||
### Clinical Investigation Workflow
|
||||
|
||||
Conduct clinical investigation:
|
||||
|
||||
1. Develop Clinical Investigation Plan (CIP)
|
||||
2. Submit to Ethics Committee for approval
|
||||
3. Notify Competent Authority (via EUDAMED)
|
||||
4. Conduct investigation per GCP principles
|
||||
5. Collect and analyze clinical data
|
||||
6. Prepare Clinical Investigation Report
|
||||
7. Submit serious adverse event reports (within 7-15 days)
|
||||
8. **Validation:** All subjects completed, data lock achieved
|
||||
|
||||
### Clinical Investigation Plan Elements
|
||||
|
||||
| Section | Content |
|
||||
|---------|---------|
|
||||
| Objectives | Primary and secondary endpoints |
|
||||
| Design | Randomized, controlled, blinded |
|
||||
| Population | Inclusion/exclusion criteria |
|
||||
| Sample size | Statistical justification |
|
||||
| Procedures | Visit schedule, assessments |
|
||||
| Endpoints | Safety and performance measures |
|
||||
| Analysis | Statistical methods |
|
||||
| Safety | Adverse event definitions, reporting |
|
||||
|
||||
### Ethics Committee Submission
|
||||
|
||||
Required documentation:
|
||||
|
||||
- Clinical Investigation Plan
|
||||
- Investigator's Brochure
|
||||
- Informed consent documents
|
||||
- Case Report Forms
|
||||
- Investigator CVs
|
||||
- Insurance certificate
|
||||
- Device documentation
|
||||
|
||||
---
|
||||
|
||||
## Post-Market Clinical Follow-up
|
||||
|
||||
### PMCF Plan Requirements
|
||||
|
||||
Develop PMCF Plan per Annex XIV Part B:
|
||||
|
||||
1. Identify residual risks from clinical evaluation
|
||||
2. Define clinical questions to address
|
||||
3. Select PMCF methods (survey, registry, study)
|
||||
4. Specify endpoints and success criteria
|
||||
5. Define timeline and milestones
|
||||
6. Plan data collection and analysis
|
||||
7. Schedule PMCF Evaluation Report updates
|
||||
8. **Validation:** PMCF Plan approved by Notified Body
|
||||
|
||||
### PMCF Methods
|
||||
|
||||
| Method | Description | Best For |
|
||||
|--------|-------------|----------|
|
||||
| Literature review | Ongoing systematic search | Mature devices |
|
||||
| Survey | User/patient questionnaire | Real-world experience |
|
||||
| Registry | Multi-site data collection | Long-term outcomes |
|
||||
| PMCF study | Prospective clinical study | Specific questions |
|
||||
| Complaint analysis | Structured complaint review | Safety signals |
|
||||
| Vigilance data | MAUDE, EUDAMED analysis | Comparative safety |
|
||||
|
||||
### PMCF Evaluation Report
|
||||
|
||||
Update frequency:
|
||||
|
||||
| Device Class | Update Frequency |
|
||||
|--------------|------------------|
|
||||
| Class III | Annual |
|
||||
| Class IIb implantable | Annual |
|
||||
| Class IIb | Every 2 years |
|
||||
| Class IIa | Every 2-5 years |
|
||||
| Class I | When clinically relevant |
|
||||
|
||||
### PMCF Report Structure
|
||||
|
||||
```
|
||||
PMCF EVALUATION REPORT
|
||||
├── 1. Executive Summary
|
||||
│ └── Key findings and conclusions
|
||||
├── 2. Scope
|
||||
│ ├── Device covered
|
||||
│ └── Reporting period
|
||||
├── 3. PMCF Activities
|
||||
│ ├── Methods employed
|
||||
│ └── Data sources
|
||||
├── 4. Results
|
||||
│ ├── Safety data
|
||||
│ ├── Performance data
|
||||
│ └── Clinical questions addressed
|
||||
├── 5. Conclusions
|
||||
│ ├── Benefit-risk confirmation
|
||||
│ ├── Residual risks updated
|
||||
│ └── Need for corrective action
|
||||
└── 6. Next Steps
|
||||
├── CER update requirements
|
||||
└── PMCF Plan modifications
|
||||
```
|
||||
|
||||
### Integration with CER
|
||||
|
||||
PMCF data feeds clinical evaluation:
|
||||
|
||||
1. PMCF data collected per plan
|
||||
2. Data analyzed in PMCF Evaluation Report
|
||||
3. CER updated with PMCF conclusions
|
||||
4. Risk management file updated
|
||||
5. IFU updated if needed
|
||||
6. **Validation:** CER update cycle completed
|
||||
@@ -0,0 +1,289 @@
|
||||
# MDR Device Classification Guide
|
||||
|
||||
EU MDR 2017/745 Annex VIII classification rules and decision framework.
|
||||
|
||||
---
|
||||
|
||||
## Table of Contents
|
||||
|
||||
- [Classification Overview](#classification-overview)
|
||||
- [Classification Rules](#classification-rules)
|
||||
- [Software Classification (MDCG 2019-11)](#software-classification)
|
||||
- [Classification Examples](#classification-examples)
|
||||
- [Conformity Assessment Routes](#conformity-assessment-routes)
|
||||
|
||||
---
|
||||
|
||||
## Classification Overview
|
||||
|
||||
### Risk Class Hierarchy
|
||||
|
||||
| Class | Risk Level | Examples | NB Required |
|
||||
|-------|------------|----------|-------------|
|
||||
| I | Lowest | Bandages, wheelchairs, stethoscopes | No (self-certification) |
|
||||
| IIa | Low-Medium | Hearing aids, dental filling materials | Yes |
|
||||
| IIb | Medium-High | Ventilators, blood bags, implantable sutures | Yes |
|
||||
| III | Highest | Pacemakers, heart valves, hip implants | Yes |
|
||||
|
||||
### Classification Factors
|
||||
|
||||
Determine class based on:
|
||||
|
||||
1. **Duration of contact:**
|
||||
- Transient: < 60 minutes
|
||||
- Short-term: 60 min to 30 days
|
||||
- Long-term: > 30 days
|
||||
|
||||
2. **Degree of invasiveness:**
|
||||
- Non-invasive
|
||||
- Invasive via body orifice
|
||||
- Surgically invasive
|
||||
- Implantable
|
||||
|
||||
3. **Body system interaction:**
|
||||
- Central circulatory system
|
||||
- Central nervous system
|
||||
- Other organ systems
|
||||
|
||||
4. **Active vs. passive:**
|
||||
- Active devices (energy dependent)
|
||||
- Passive devices
|
||||
|
||||
---
|
||||
|
||||
## Classification Rules
|
||||
|
||||
### Non-Invasive Devices (Rules 1-4)
|
||||
|
||||
**Rule 1 - General non-invasive:**
|
||||
- Class I (unless covered by other rules)
|
||||
- Example: Wheelchairs, hospital beds, collection devices
|
||||
|
||||
**Rule 2 - Channeling or storing:**
|
||||
- Class IIa: Blood bags, transfusion sets (>60 min contact)
|
||||
- Class IIb: Blood storage, organ storage
|
||||
- Class I: Simple channeling (gravity, IV bag without additives)
|
||||
|
||||
**Rule 3 - Modifying biological composition:**
|
||||
- Class IIa: Filters, gas separators, dialysis filters
|
||||
- Class IIb: Blood filtration, exchange transfusion
|
||||
|
||||
**Rule 4 - Contact with injured skin:**
|
||||
- Class I: Wound dressings for superficial wounds
|
||||
- Class IIa: Wounds in dermis requiring secondary intent healing
|
||||
- Class IIb: Severe wounds, chronic wounds, burns
|
||||
|
||||
### Invasive Devices (Rules 5-8)
|
||||
|
||||
**Rule 5 - Body orifice invasive (transient):**
|
||||
- Class I: Transient use, non-surgically invasive
|
||||
- Class IIa: Short-term use
|
||||
- Class IIb: Long-term use in oral cavity
|
||||
|
||||
**Rule 6 - Surgically invasive (transient):**
|
||||
- Class IIa: Transient use
|
||||
- Exception Class I: Reusable surgical instruments
|
||||
|
||||
**Rule 7 - Surgically invasive (short-term):**
|
||||
- Class IIa: Short-term (< 30 days)
|
||||
- Class IIb: Central circulatory or CNS contact
|
||||
- Class III: Chemical change or drug delivery
|
||||
|
||||
**Rule 8 - Implantable and long-term surgically invasive:**
|
||||
- Class IIb: General implants
|
||||
- Class III: Heart, CNS, spine contact; drug delivery; biological origin
|
||||
|
||||
### Active Devices (Rules 9-13)
|
||||
|
||||
**Rule 9 - Active therapeutic devices:**
|
||||
- Class IIa: Exchange or admin of energy (non-hazardous)
|
||||
- Class IIb: Potentially hazardous energy levels
|
||||
|
||||
**Rule 10 - Active diagnostic devices:**
|
||||
- Class IIa: Supply energy for imaging, monitoring
|
||||
- Class IIb: Monitor vital physiological parameters
|
||||
|
||||
**Rule 11 - Software:**
|
||||
- Class IIa: Information for diagnostic/therapeutic decisions (non-serious)
|
||||
- Class IIb: Decisions that could cause death/irreversible deterioration
|
||||
- Class III: Decisions with immediate risk to life
|
||||
- See MDCG 2019-11 for detailed algorithm
|
||||
|
||||
**Rule 12 - Active devices administering substances:**
|
||||
- Class IIa: Non-hazardous manner
|
||||
- Class IIb: Potentially hazardous manner
|
||||
|
||||
**Rule 13 - Other active devices:**
|
||||
- Class I: All other active devices
|
||||
|
||||
### Special Rules (Rules 14-22)
|
||||
|
||||
**Rule 14 - Contraception/STI prevention:**
|
||||
- Class IIb: Contraceptive devices
|
||||
- Class III: Implantable contraceptives
|
||||
|
||||
**Rule 15 - Disinfection/sterilization:**
|
||||
- Class IIa: Disinfection of devices
|
||||
- Class IIb: Disinfection of invasive devices
|
||||
|
||||
**Rule 16 - X-ray diagnostic recording:**
|
||||
- Class IIa: Recording media for x-ray
|
||||
|
||||
**Rule 17 - Devices with nanomaterials:**
|
||||
- Class III: High internal exposure potential
|
||||
- Class IIb: Medium exposure
|
||||
- Class IIa: Low exposure
|
||||
|
||||
**Rule 18 - Blood/plasma derivatives:**
|
||||
- Class III: Utilizing blood derivatives
|
||||
|
||||
**Rule 19 - Drug delivery systems:**
|
||||
- Class III: Integral drug administration
|
||||
|
||||
**Rule 20 - Breath analyzers for anesthesia:**
|
||||
- Class IIb: Breath analyzers
|
||||
|
||||
**Rule 21 - Medicinal substance devices:**
|
||||
- Class III: Incorporating medicinal substances
|
||||
|
||||
**Rule 22 - Closed-loop therapeutic systems:**
|
||||
- Class III: Closed-loop systems
|
||||
|
||||
---
|
||||
|
||||
## Software Classification
|
||||
|
||||
### MDCG 2019-11 Decision Algorithm
|
||||
|
||||
Execute software classification:
|
||||
|
||||
1. Determine if software qualifies as medical device
|
||||
2. Identify significance of information to healthcare decision
|
||||
3. Assess healthcare situation or patient condition
|
||||
4. Apply rule 11 based on severity
|
||||
5. **Validation:** Classification rationale documented with MDCG reference
|
||||
|
||||
### Software Classification Matrix
|
||||
|
||||
| Information Significance | Situation/Condition | Class |
|
||||
|--------------------------|---------------------|-------|
|
||||
| Informs clinical management | Non-serious | IIa |
|
||||
| Informs clinical management | Serious | IIb |
|
||||
| Informs clinical management | Critical | III |
|
||||
| Drives clinical management | Non-serious | IIa |
|
||||
| Drives clinical management | Serious | IIb |
|
||||
| Drives clinical management | Critical | III |
|
||||
| Treats or diagnoses | Non-serious | IIa |
|
||||
| Treats or diagnoses | Serious | IIb |
|
||||
| Treats or diagnoses | Critical | III |
|
||||
|
||||
### Software Examples
|
||||
|
||||
| Software Type | Class | Rationale |
|
||||
|---------------|-------|-----------|
|
||||
| Patient record viewing | Not MD | Administrative, not clinical |
|
||||
| Medication reminder app | Class I | General wellness |
|
||||
| Blood glucose monitor app | Class IIa | Informs non-serious decisions |
|
||||
| Sepsis detection algorithm | Class IIb | Informs serious condition |
|
||||
| AI tumor detection | Class III | Diagnoses critical condition |
|
||||
| Closed-loop insulin delivery | Class III | Treats critical condition |
|
||||
|
||||
---
|
||||
|
||||
## Classification Examples
|
||||
|
||||
### Example 1: Surgical Suture (Absorbable)
|
||||
|
||||
```
|
||||
Device: Absorbable suture for internal wound closure
|
||||
Analysis:
|
||||
- Invasiveness: Surgically invasive
|
||||
- Duration: Long-term (absorbed over > 30 days)
|
||||
- System: General tissue (not CNS, not cardiac)
|
||||
- Rule Applied: Rule 8 (implantable, long-term)
|
||||
|
||||
Classification: Class IIb
|
||||
Rationale: Implantable device > 30 days, general tissue
|
||||
Conformity Route: Annex IX (Type examination) + Annex XI
|
||||
```
|
||||
|
||||
### Example 2: Blood Pressure Monitor
|
||||
|
||||
```
|
||||
Device: Home blood pressure monitoring device
|
||||
Analysis:
|
||||
- Active: Yes (electronic measurement)
|
||||
- Function: Monitoring vital physiological parameter
|
||||
- Risk: Non-immediate (home use, not ICU)
|
||||
- Rule Applied: Rule 10 (active diagnostic)
|
||||
|
||||
Classification: Class IIa
|
||||
Rationale: Monitors vital parameter, non-critical setting
|
||||
Conformity Route: Annex IX or XI (QMS + product verification)
|
||||
```
|
||||
|
||||
### Example 3: Hip Implant
|
||||
|
||||
```
|
||||
Device: Total hip replacement prosthesis
|
||||
Analysis:
|
||||
- Invasiveness: Surgically invasive, implantable
|
||||
- Duration: Long-term (permanent)
|
||||
- System: Musculoskeletal
|
||||
- Rule Applied: Rule 8 (implantable, long-term)
|
||||
|
||||
Classification: Class III
|
||||
Rationale: Implantable > 30 days in direct contact with bone
|
||||
Conformity Route: Annex IX + Annex X (full QMS + type examination)
|
||||
```
|
||||
|
||||
### Example 4: Diagnostic Software (AI)
|
||||
|
||||
```
|
||||
Device: AI-based chest X-ray analysis for pneumonia detection
|
||||
Analysis:
|
||||
- Software: Qualifies as medical device (clinical decision)
|
||||
- Information: Diagnoses condition
|
||||
- Condition: Serious (pneumonia can be life-threatening)
|
||||
- Rule Applied: Rule 11 + MDCG 2019-11
|
||||
|
||||
Classification: Class IIb
|
||||
Rationale: Software diagnosing serious condition
|
||||
Conformity Route: Annex IX or Annex XI
|
||||
```
|
||||
|
||||
---
|
||||
|
||||
## Conformity Assessment Routes
|
||||
|
||||
### By Device Class
|
||||
|
||||
| Class | Conformity Route | NB Involvement |
|
||||
|-------|------------------|----------------|
|
||||
| I | Annex II (self-declaration) | None |
|
||||
| I (sterile/measuring) | Annex II + IX/XI | Sterile/measuring aspects |
|
||||
| IIa | Annex II + IX or XI | Product verification or QMS |
|
||||
| IIb | Annex IX + X or Annex X + XI | Type exam + QMS or production |
|
||||
| III | Annex IX + X | Full QMS + type examination |
|
||||
|
||||
### Annex Reference
|
||||
|
||||
| Annex | Content | Purpose |
|
||||
|-------|---------|---------|
|
||||
| II | Technical documentation | Required for all classes |
|
||||
| III | Technical documentation (additions) | Class III additions |
|
||||
| IX | Conformity assessment (QMS) | Quality management route |
|
||||
| X | Type examination | Product design examination |
|
||||
| XI | Product verification | Production quality checks |
|
||||
|
||||
### Decision Workflow
|
||||
|
||||
Select conformity route:
|
||||
|
||||
1. Determine device classification (Rules 1-22)
|
||||
2. Identify applicable annexes for class
|
||||
3. Evaluate QMS maturity (Annex IX capability)
|
||||
4. Consider production volume (batch vs. mass)
|
||||
5. Assess Notified Body capacity and timeline
|
||||
6. Select optimal conformity assessment route
|
||||
7. **Validation:** Route confirmed with Notified Body consultation
|
||||
@@ -0,0 +1,355 @@
|
||||
# Technical Documentation Templates
|
||||
|
||||
MDR Annex II and III technical file structure and content requirements.
|
||||
|
||||
---
|
||||
|
||||
## Table of Contents
|
||||
|
||||
- [Technical Documentation Overview](#technical-documentation-overview)
|
||||
- [Annex II Requirements](#annex-ii-requirements)
|
||||
- [Annex III Additions](#annex-iii-additions)
|
||||
- [Document Templates](#document-templates)
|
||||
- [Notified Body Expectations](#notified-body-expectations)
|
||||
|
||||
---
|
||||
|
||||
## Technical Documentation Overview
|
||||
|
||||
### Documentation Hierarchy
|
||||
|
||||
```
|
||||
TECHNICAL DOCUMENTATION
|
||||
├── Device Description and Specification
|
||||
├── Information Supplied by Manufacturer
|
||||
├── Design and Manufacturing Information
|
||||
├── General Safety and Performance Requirements
|
||||
├── Benefit-Risk Analysis
|
||||
├── Product Verification and Validation
|
||||
├── Clinical Evaluation Report
|
||||
└── Post-Market Surveillance Documentation
|
||||
```
|
||||
|
||||
### Documentation by Phase
|
||||
|
||||
| Phase | Required Documents |
|
||||
|-------|-------------------|
|
||||
| Design Input | User needs, design requirements, regulatory requirements |
|
||||
| Design Development | Design specifications, drawings, BOM, software docs |
|
||||
| Verification | Test protocols, test reports, design review records |
|
||||
| Validation | Clinical data, usability data, biocompatibility |
|
||||
| Transfer | Manufacturing specs, process validations |
|
||||
| Post-Market | PMS plan, PMCF plan, vigilance procedures |
|
||||
|
||||
---
|
||||
|
||||
## Annex II Requirements
|
||||
|
||||
### Section 1: Device Description and Specification
|
||||
|
||||
**1.1 Device Identification**
|
||||
|
||||
```
|
||||
DEVICE IDENTIFICATION
|
||||
├── Trade name(s)
|
||||
├── General description of the device
|
||||
├── Basic UDI-DI
|
||||
├── Device identifier codes (internal + regulatory)
|
||||
├── Intended purpose statement
|
||||
├── Indications for use
|
||||
├── Contraindications
|
||||
├── Target population (patient, user)
|
||||
├── Medical conditions intended to diagnose/treat
|
||||
└── Principles of operation
|
||||
```
|
||||
|
||||
**1.2 Device Variants and Accessories**
|
||||
|
||||
| Element | Description |
|
||||
|---------|-------------|
|
||||
| Variant listing | All variants with identifiers |
|
||||
| Configuration differences | Technical differences by variant |
|
||||
| Accessories | Separate devices used together |
|
||||
| Spare parts | Replaceable components |
|
||||
|
||||
**1.3 Reference to Previous Generations**
|
||||
|
||||
- Previous generation device identification
|
||||
- Key modifications summary
|
||||
- Clinical experience from prior device
|
||||
- Justification for changes
|
||||
|
||||
### Section 2: Information Supplied by Manufacturer
|
||||
|
||||
**2.1 Label Requirements**
|
||||
|
||||
Mandatory label elements per Article 13:
|
||||
|
||||
- [ ] Device name or trade name
|
||||
- [ ] Manufacturer name and address
|
||||
- [ ] Authorized representative (if applicable)
|
||||
- [ ] Lot/batch number or serial number
|
||||
- [ ] UDI carrier (AIDC + HRI)
|
||||
- [ ] Expiration date (if applicable)
|
||||
- [ ] Storage/handling conditions
|
||||
- [ ] Warnings and precautions
|
||||
- [ ] CE mark with NB number (if applicable)
|
||||
- [ ] Symbol meanings per EN ISO 15223-1
|
||||
|
||||
**2.2 Instructions for Use**
|
||||
|
||||
IFU structure:
|
||||
|
||||
```
|
||||
INSTRUCTIONS FOR USE
|
||||
├── 1. Device Description
|
||||
│ ├── Intended purpose
|
||||
│ ├── Indications and contraindications
|
||||
│ └── Principle of operation
|
||||
├── 2. Warnings and Precautions
|
||||
│ ├── Contraindicated uses
|
||||
│ ├── Potential complications
|
||||
│ └── Drug/device interactions
|
||||
├── 3. User Instructions
|
||||
│ ├── Unpacking and inspection
|
||||
│ ├── Setup/installation
|
||||
│ ├── Operating procedures
|
||||
│ └── Cleaning/maintenance
|
||||
├── 4. Technical Specifications
|
||||
│ ├── Physical characteristics
|
||||
│ ├── Performance characteristics
|
||||
│ └── Environmental limits
|
||||
├── 5. Troubleshooting
|
||||
│ ├── Error codes/messages
|
||||
│ └── Corrective actions
|
||||
└── 6. Symbols Glossary
|
||||
```
|
||||
|
||||
### Section 3: Design and Manufacturing Information
|
||||
|
||||
**3.1 Design Process Documentation**
|
||||
|
||||
| Document | Purpose |
|
||||
|----------|---------|
|
||||
| Design input | User needs, regulatory requirements |
|
||||
| Design output | Specifications, drawings, software |
|
||||
| Design review | Review records at key milestones |
|
||||
| Design verification | Test protocols and results |
|
||||
| Design validation | Clinical/usability evidence |
|
||||
| Design transfer | Manufacturing readiness |
|
||||
| Design changes | Change control records |
|
||||
|
||||
**3.2 Manufacturing Process Description**
|
||||
|
||||
```
|
||||
MANUFACTURING DOCUMENTATION
|
||||
├── Process flow diagram
|
||||
├── Manufacturing specifications
|
||||
├── Facility and equipment qualification
|
||||
├── Process validation protocols/reports
|
||||
├── Environmental monitoring
|
||||
├── Personnel training records
|
||||
├── In-process controls
|
||||
├── Final inspection/testing
|
||||
├── Sterilization validation (if applicable)
|
||||
└── Packaging validation
|
||||
```
|
||||
|
||||
**3.3 Supplier and Subcontractor Information**
|
||||
|
||||
- Approved supplier list
|
||||
- Supplier qualification records
|
||||
- Critical component specifications
|
||||
- Incoming inspection procedures
|
||||
- Supplier audit records
|
||||
|
||||
### Section 4: General Safety and Performance Requirements
|
||||
|
||||
**GSPR Compliance Checklist**
|
||||
|
||||
| GSPR | Requirement | Evidence |
|
||||
|------|-------------|----------|
|
||||
| 1 | Safe design for intended use | Risk management file |
|
||||
| 2 | Risk acceptable when weighed against benefits | Benefit-risk analysis |
|
||||
| 3 | State of the art design | Literature review, standards |
|
||||
| 4 | No compromise of clinical condition | Clinical evaluation |
|
||||
| 5 | Transport and storage conditions | Shelf life testing |
|
||||
| 6 | Acceptable undesirable effects | Risk-benefit analysis |
|
||||
| 7 | CE marking conformity | Declaration of conformity |
|
||||
| ... | Continue for all applicable GSPRs | |
|
||||
|
||||
**GSPR Matrix Template**
|
||||
|
||||
| GSPR # | Requirement Summary | Applicable? | Evidence Document | Status |
|
||||
|--------|---------------------|-------------|-------------------|--------|
|
||||
| 10.1 | Chemical properties | Yes/No/NA | Biocompatibility report | Complete |
|
||||
| 10.2 | Infection risk | Yes/No/NA | Sterilization validation | Complete |
|
||||
| 10.3 | Substances with carcinogenic risk | Yes/No/NA | Material specification | Complete |
|
||||
|
||||
### Section 5: Benefit-Risk Analysis
|
||||
|
||||
**Benefit-Risk Documentation**
|
||||
|
||||
```
|
||||
BENEFIT-RISK ANALYSIS
|
||||
├── 1. Intended Benefits
|
||||
│ ├── Direct therapeutic benefits
|
||||
│ ├── Diagnostic accuracy improvements
|
||||
│ └── Patient outcome benefits
|
||||
├── 2. Known Risks
|
||||
│ ├── Identified hazards (from risk analysis)
|
||||
│ ├── Risk control measures implemented
|
||||
│ └── Residual risks
|
||||
├── 3. Benefit-Risk Determination
|
||||
│ ├── Qualitative analysis
|
||||
│ ├── Quantitative analysis (if available)
|
||||
│ └── Comparison to alternatives
|
||||
└── 4. Conclusion
|
||||
├── Acceptability statement
|
||||
└── Justification for residual risks
|
||||
```
|
||||
|
||||
### Section 6: Product Verification and Validation
|
||||
|
||||
**6.1 Verification Testing**
|
||||
|
||||
| Test Category | Standards | Documentation |
|
||||
|---------------|-----------|---------------|
|
||||
| Electrical safety | IEC 60601-1 | Test protocol + report |
|
||||
| EMC | IEC 60601-1-2 | EMC test report |
|
||||
| Biocompatibility | ISO 10993 series | Biocompatibility evaluation |
|
||||
| Software | IEC 62304 | Software verification report |
|
||||
| Sterilization | ISO 11135/11137 | Sterility assurance |
|
||||
| Packaging | ISO 11607 | Packaging validation |
|
||||
| Shelf life | Accelerated aging | Stability study report |
|
||||
| Usability | IEC 62366-1 | Usability engineering file |
|
||||
|
||||
**6.2 Validation Evidence**
|
||||
|
||||
- Clinical investigation data
|
||||
- Literature-based clinical evidence
|
||||
- Simulated use testing
|
||||
- User feedback/complaint analysis
|
||||
- Post-market surveillance data
|
||||
|
||||
---
|
||||
|
||||
## Annex III Additions
|
||||
|
||||
### Class III Specific Requirements
|
||||
|
||||
Additional documentation for Class III devices:
|
||||
|
||||
**Implant-Specific Requirements**
|
||||
|
||||
- Implant card information
|
||||
- Patient information leaflet
|
||||
- Device tracking procedures
|
||||
- Explant analysis capability
|
||||
|
||||
**Drug-Device Combination**
|
||||
|
||||
- Drug substance specification
|
||||
- Drug compatibility testing
|
||||
- Combined product assessment
|
||||
- Pharmacovigilance interface
|
||||
|
||||
---
|
||||
|
||||
## Document Templates
|
||||
|
||||
### Design History File Index
|
||||
|
||||
```
|
||||
DESIGN HISTORY FILE (DHF)
|
||||
Document ID: DHF-[Product]-[Rev]
|
||||
|
||||
1. DESIGN INPUT
|
||||
1.1 User Requirements Specification (URS)
|
||||
1.2 Regulatory Requirements Matrix
|
||||
1.3 Design Input Review Record
|
||||
|
||||
2. DESIGN OUTPUT
|
||||
2.1 Product Specification
|
||||
2.2 Engineering Drawings
|
||||
2.3 Bill of Materials
|
||||
2.4 Software Documentation
|
||||
|
||||
3. DESIGN VERIFICATION
|
||||
3.1 Verification Test Plan
|
||||
3.2 Verification Test Reports
|
||||
3.3 Traceability Matrix
|
||||
|
||||
4. DESIGN VALIDATION
|
||||
4.1 Clinical Evaluation Report
|
||||
4.2 Usability Engineering File
|
||||
4.3 Biocompatibility Evaluation
|
||||
|
||||
5. DESIGN TRANSFER
|
||||
5.1 Manufacturing Procedures
|
||||
5.2 Process Validation Reports
|
||||
5.3 Supplier Qualification
|
||||
|
||||
6. DESIGN REVIEWS
|
||||
6.1 Design Review Records
|
||||
6.2 Risk Management Review
|
||||
6.3 Final Design Release
|
||||
```
|
||||
|
||||
### Declaration of Conformity Template
|
||||
|
||||
```
|
||||
EU DECLARATION OF CONFORMITY
|
||||
|
||||
We, [Manufacturer Name]
|
||||
Address: [Full address]
|
||||
|
||||
declare under our sole responsibility that the device:
|
||||
|
||||
Device name: [Trade name]
|
||||
Device description: [Description]
|
||||
Basic UDI-DI: [UDI-DI]
|
||||
Classification: [Class I/IIa/IIb/III]
|
||||
|
||||
is in conformity with the provisions of:
|
||||
- Regulation (EU) 2017/745
|
||||
|
||||
Applicable standards:
|
||||
- [List harmonized standards]
|
||||
|
||||
Notified Body: [NB name and number] (if applicable)
|
||||
Certificate number: [Certificate number]
|
||||
|
||||
Place and date: [Location, Date]
|
||||
Signature: [Authorized signatory]
|
||||
Name and function: [Name, Title]
|
||||
```
|
||||
|
||||
---
|
||||
|
||||
## Notified Body Expectations
|
||||
|
||||
### Common NB Findings
|
||||
|
||||
| Finding Area | Common Issue | Prevention |
|
||||
|--------------|--------------|------------|
|
||||
| GSPR matrix | Incomplete, no evidence links | Complete matrix with references |
|
||||
| Risk management | Not integrated with design | Update throughout development |
|
||||
| Clinical evaluation | Insufficient literature search | Systematic search with PICO |
|
||||
| IFU | Missing warnings | Risk-based IFU content |
|
||||
| Traceability | Design to requirements gaps | Maintain traceability matrix |
|
||||
|
||||
### Pre-Submission Checklist
|
||||
|
||||
Before Notified Body submission:
|
||||
|
||||
- [ ] Technical documentation complete
|
||||
- [ ] GSPR checklist fully addressed
|
||||
- [ ] Risk management file current
|
||||
- [ ] Clinical evaluation report complete
|
||||
- [ ] QMS documentation ready
|
||||
- [ ] Design verification complete
|
||||
- [ ] Design validation complete
|
||||
- [ ] Labeling and IFU finalized
|
||||
- [ ] Declaration of conformity prepared
|
||||
- [ ] **Validation:** Internal review completed
|
||||
@@ -1,19 +0,0 @@
|
||||
#!/usr/bin/env python3
|
||||
"""
|
||||
Example helper script for mdr-745-specialist
|
||||
|
||||
This is a placeholder script that can be executed directly.
|
||||
Replace with actual implementation or delete if not needed.
|
||||
|
||||
Example real scripts from other skills:
|
||||
- pdf/scripts/fill_fillable_fields.py - Fills PDF form fields
|
||||
- pdf/scripts/convert_pdf_to_images.py - Converts PDF pages to images
|
||||
"""
|
||||
|
||||
def main():
|
||||
print("This is an example script for mdr-745-specialist")
|
||||
# TODO: Add actual script logic here
|
||||
# This could be data processing, file conversion, API calls, etc.
|
||||
|
||||
if __name__ == "__main__":
|
||||
main()
|
||||
511
ra-qm-team/mdr-745-specialist/scripts/mdr_gap_analyzer.py
Normal file
511
ra-qm-team/mdr-745-specialist/scripts/mdr_gap_analyzer.py
Normal file
@@ -0,0 +1,511 @@
|
||||
#!/usr/bin/env python3
|
||||
"""
|
||||
MDR Gap Analyzer - EU MDR 2017/745 Compliance Gap Assessment Tool
|
||||
|
||||
Analyzes device classification, identifies documentation gaps, and generates
|
||||
compliance roadmap for EU MDR transition.
|
||||
|
||||
Usage:
|
||||
python mdr_gap_analyzer.py --device "Device Name" --class IIa
|
||||
python mdr_gap_analyzer.py --device "Device Name" --class III --output json
|
||||
python mdr_gap_analyzer.py --interactive
|
||||
"""
|
||||
|
||||
import argparse
|
||||
import json
|
||||
import sys
|
||||
from dataclasses import dataclass, field, asdict
|
||||
from datetime import datetime
|
||||
from typing import List, Dict, Optional
|
||||
from enum import Enum
|
||||
|
||||
|
||||
class DeviceClass(Enum):
|
||||
I = "I"
|
||||
I_STERILE = "Is"
|
||||
I_MEASURING = "Im"
|
||||
IIA = "IIa"
|
||||
IIB = "IIb"
|
||||
III = "III"
|
||||
|
||||
|
||||
class GapStatus(Enum):
|
||||
NOT_STARTED = "Not Started"
|
||||
IN_PROGRESS = "In Progress"
|
||||
COMPLETE = "Complete"
|
||||
NOT_APPLICABLE = "N/A"
|
||||
|
||||
|
||||
@dataclass
|
||||
class GapItem:
|
||||
requirement: str
|
||||
category: str
|
||||
description: str
|
||||
status: GapStatus = GapStatus.NOT_STARTED
|
||||
priority: str = "Medium"
|
||||
evidence_needed: List[str] = field(default_factory=list)
|
||||
notes: str = ""
|
||||
|
||||
|
||||
@dataclass
|
||||
class GapAnalysisResult:
|
||||
device_name: str
|
||||
device_class: str
|
||||
analysis_date: str
|
||||
total_requirements: int
|
||||
gaps_identified: int
|
||||
completion_percentage: float
|
||||
gaps: List[Dict]
|
||||
recommendations: List[str]
|
||||
critical_gaps: List[str]
|
||||
|
||||
|
||||
class MDRGapAnalyzer:
|
||||
"""Analyzer for EU MDR 2017/745 compliance gaps."""
|
||||
|
||||
# MDR Requirements by category
|
||||
REQUIREMENTS = {
|
||||
"technical_documentation": [
|
||||
GapItem(
|
||||
requirement="Annex II - Device Description",
|
||||
category="Technical Documentation",
|
||||
description="Complete device description including variants, accessories, intended purpose",
|
||||
priority="High",
|
||||
evidence_needed=["Device specification", "Intended purpose statement", "Variant listing"]
|
||||
),
|
||||
GapItem(
|
||||
requirement="Annex II - Information Supplied",
|
||||
category="Technical Documentation",
|
||||
description="Label and IFU meeting Article 13 requirements",
|
||||
priority="High",
|
||||
evidence_needed=["Label artwork", "Instructions for use", "Symbol glossary"]
|
||||
),
|
||||
GapItem(
|
||||
requirement="Annex II - Design and Manufacturing",
|
||||
category="Technical Documentation",
|
||||
description="Design history file and manufacturing documentation",
|
||||
priority="High",
|
||||
evidence_needed=["Design history file", "Process flow diagram", "Validation reports"]
|
||||
),
|
||||
GapItem(
|
||||
requirement="Annex II - GSPR Compliance",
|
||||
category="Technical Documentation",
|
||||
description="General Safety and Performance Requirements checklist",
|
||||
priority="Critical",
|
||||
evidence_needed=["GSPR matrix", "Standard compliance evidence", "Risk management file"]
|
||||
),
|
||||
],
|
||||
"clinical_evaluation": [
|
||||
GapItem(
|
||||
requirement="Annex XIV Part A - Clinical Evaluation",
|
||||
category="Clinical Evaluation",
|
||||
description="Clinical evaluation report with systematic literature review",
|
||||
priority="Critical",
|
||||
evidence_needed=["Clinical evaluation report", "Literature search protocol", "Data appraisal"]
|
||||
),
|
||||
GapItem(
|
||||
requirement="Annex XIV Part B - PMCF",
|
||||
category="Clinical Evaluation",
|
||||
description="Post-market clinical follow-up plan and evaluation report",
|
||||
priority="High",
|
||||
evidence_needed=["PMCF plan", "PMCF evaluation report", "Residual risk assessment"]
|
||||
),
|
||||
GapItem(
|
||||
requirement="Qualified Person for CER",
|
||||
category="Clinical Evaluation",
|
||||
description="Clinical evaluation by qualified evaluator per Annex XIV",
|
||||
priority="High",
|
||||
evidence_needed=["Evaluator CV", "Qualification evidence", "Signed CER"]
|
||||
),
|
||||
],
|
||||
"risk_management": [
|
||||
GapItem(
|
||||
requirement="ISO 14971 Risk Management",
|
||||
category="Risk Management",
|
||||
description="Complete risk management file per ISO 14971:2019",
|
||||
priority="Critical",
|
||||
evidence_needed=["Risk management plan", "Risk analysis", "Risk evaluation", "Risk control"]
|
||||
),
|
||||
GapItem(
|
||||
requirement="Benefit-Risk Analysis",
|
||||
category="Risk Management",
|
||||
description="Documented benefit-risk determination",
|
||||
priority="High",
|
||||
evidence_needed=["Benefit-risk analysis document", "Residual risk acceptability"]
|
||||
),
|
||||
],
|
||||
"quality_management": [
|
||||
GapItem(
|
||||
requirement="ISO 13485 QMS",
|
||||
category="Quality Management",
|
||||
description="Quality management system conforming to ISO 13485:2016",
|
||||
priority="Critical",
|
||||
evidence_needed=["QMS manual", "Process documentation", "Internal audit records"]
|
||||
),
|
||||
GapItem(
|
||||
requirement="Post-Market Surveillance",
|
||||
category="Quality Management",
|
||||
description="PMS system per Article 83-86",
|
||||
priority="High",
|
||||
evidence_needed=["PMS plan", "PSUR (if required)", "Vigilance procedures"]
|
||||
),
|
||||
],
|
||||
"udi_eudamed": [
|
||||
GapItem(
|
||||
requirement="UDI System",
|
||||
category="UDI/EUDAMED",
|
||||
description="Unique Device Identification per Article 27",
|
||||
priority="High",
|
||||
evidence_needed=["UDI-DI assignment", "Label with UDI carrier", "GUDID/EUDAMED registration"]
|
||||
),
|
||||
GapItem(
|
||||
requirement="EUDAMED Registration",
|
||||
category="UDI/EUDAMED",
|
||||
description="Actor, device, and certificate registration in EUDAMED",
|
||||
priority="Medium",
|
||||
evidence_needed=["Actor registration", "Device registration", "Certificate upload"]
|
||||
),
|
||||
],
|
||||
"notified_body": [
|
||||
GapItem(
|
||||
requirement="Notified Body Selection",
|
||||
category="Notified Body",
|
||||
description="Selection and engagement of MDR-designated Notified Body",
|
||||
priority="Critical",
|
||||
evidence_needed=["NB selection criteria", "NB engagement letter", "Audit schedule"]
|
||||
),
|
||||
GapItem(
|
||||
requirement="Conformity Assessment",
|
||||
category="Notified Body",
|
||||
description="Completion of appropriate conformity assessment procedure",
|
||||
priority="Critical",
|
||||
evidence_needed=["Application dossier", "Technical documentation submission", "Certificate"]
|
||||
),
|
||||
],
|
||||
}
|
||||
|
||||
# Class-specific requirements
|
||||
CLASS_REQUIREMENTS = {
|
||||
DeviceClass.III: [
|
||||
GapItem(
|
||||
requirement="Annex III - Class III Additions",
|
||||
category="Technical Documentation",
|
||||
description="Additional documentation for Class III devices",
|
||||
priority="Critical",
|
||||
evidence_needed=["Implant card", "Patient information", "Device tracking"]
|
||||
),
|
||||
GapItem(
|
||||
requirement="Clinical Investigation",
|
||||
category="Clinical Evaluation",
|
||||
description="Clinical investigation per Article 61 (unless equivalent device)",
|
||||
priority="Critical",
|
||||
evidence_needed=["Clinical investigation plan", "Ethics approval", "Clinical study report"]
|
||||
),
|
||||
],
|
||||
DeviceClass.IIB: [
|
||||
GapItem(
|
||||
requirement="Implantable Device Documentation",
|
||||
category="Technical Documentation",
|
||||
description="Additional requirements for implantable Class IIb devices",
|
||||
priority="High",
|
||||
evidence_needed=["Implant card (if implantable)", "Long-term safety data"]
|
||||
),
|
||||
],
|
||||
}
|
||||
|
||||
def __init__(self, device_name: str, device_class: DeviceClass):
|
||||
self.device_name = device_name
|
||||
self.device_class = device_class
|
||||
self.gaps: List[GapItem] = []
|
||||
self._build_requirements_list()
|
||||
|
||||
def _build_requirements_list(self):
|
||||
"""Build complete requirements list based on device class."""
|
||||
# Add all base requirements
|
||||
for category_gaps in self.REQUIREMENTS.values():
|
||||
for gap in category_gaps:
|
||||
self.gaps.append(GapItem(
|
||||
requirement=gap.requirement,
|
||||
category=gap.category,
|
||||
description=gap.description,
|
||||
priority=gap.priority,
|
||||
evidence_needed=gap.evidence_needed.copy()
|
||||
))
|
||||
|
||||
# Add class-specific requirements
|
||||
if self.device_class in self.CLASS_REQUIREMENTS:
|
||||
for gap in self.CLASS_REQUIREMENTS[self.device_class]:
|
||||
self.gaps.append(GapItem(
|
||||
requirement=gap.requirement,
|
||||
category=gap.category,
|
||||
description=gap.description,
|
||||
priority=gap.priority,
|
||||
evidence_needed=gap.evidence_needed.copy()
|
||||
))
|
||||
|
||||
# Class I self-certification: NB not required
|
||||
if self.device_class == DeviceClass.I:
|
||||
for gap in self.gaps:
|
||||
if gap.category == "Notified Body":
|
||||
gap.status = GapStatus.NOT_APPLICABLE
|
||||
|
||||
def update_gap_status(self, requirement: str, status: GapStatus, notes: str = ""):
|
||||
"""Update status of a specific gap."""
|
||||
for gap in self.gaps:
|
||||
if gap.requirement == requirement:
|
||||
gap.status = status
|
||||
gap.notes = notes
|
||||
break
|
||||
|
||||
def analyze(self) -> GapAnalysisResult:
|
||||
"""Perform gap analysis and generate results."""
|
||||
applicable_gaps = [g for g in self.gaps if g.status != GapStatus.NOT_APPLICABLE]
|
||||
complete_gaps = [g for g in applicable_gaps if g.status == GapStatus.COMPLETE]
|
||||
|
||||
completion = (len(complete_gaps) / len(applicable_gaps) * 100) if applicable_gaps else 0
|
||||
|
||||
# Identify critical gaps
|
||||
critical_gaps = [
|
||||
g.requirement for g in applicable_gaps
|
||||
if g.priority == "Critical" and g.status != GapStatus.COMPLETE
|
||||
]
|
||||
|
||||
# Generate recommendations
|
||||
recommendations = self._generate_recommendations()
|
||||
|
||||
return GapAnalysisResult(
|
||||
device_name=self.device_name,
|
||||
device_class=self.device_class.value,
|
||||
analysis_date=datetime.now().isoformat(),
|
||||
total_requirements=len(applicable_gaps),
|
||||
gaps_identified=len(applicable_gaps) - len(complete_gaps),
|
||||
completion_percentage=round(completion, 1),
|
||||
gaps=[{
|
||||
"requirement": g.requirement,
|
||||
"category": g.category,
|
||||
"status": g.status.value,
|
||||
"priority": g.priority,
|
||||
"evidence_needed": g.evidence_needed
|
||||
} for g in applicable_gaps],
|
||||
recommendations=recommendations,
|
||||
critical_gaps=critical_gaps
|
||||
)
|
||||
|
||||
def _generate_recommendations(self) -> List[str]:
|
||||
"""Generate prioritized recommendations."""
|
||||
recommendations = []
|
||||
|
||||
# Check for critical gaps
|
||||
critical_incomplete = [
|
||||
g for g in self.gaps
|
||||
if g.priority == "Critical" and g.status not in [GapStatus.COMPLETE, GapStatus.NOT_APPLICABLE]
|
||||
]
|
||||
|
||||
if critical_incomplete:
|
||||
recommendations.append(
|
||||
f"CRITICAL: {len(critical_incomplete)} critical requirements not complete. "
|
||||
"Address immediately to proceed with conformity assessment."
|
||||
)
|
||||
|
||||
# Check clinical evaluation
|
||||
cer_gap = next((g for g in self.gaps if "Clinical Evaluation" in g.requirement), None)
|
||||
if cer_gap and cer_gap.status != GapStatus.COMPLETE:
|
||||
recommendations.append(
|
||||
"Clinical Evaluation Report (CER) is incomplete. "
|
||||
"This is required before Notified Body submission."
|
||||
)
|
||||
|
||||
# Check for Class III specific
|
||||
if self.device_class == DeviceClass.III:
|
||||
ci_gap = next((g for g in self.gaps if "Clinical Investigation" in g.requirement), None)
|
||||
if ci_gap and ci_gap.status != GapStatus.COMPLETE:
|
||||
recommendations.append(
|
||||
"Class III device requires clinical investigation per Article 61 "
|
||||
"unless equivalence can be demonstrated."
|
||||
)
|
||||
|
||||
# Check EUDAMED
|
||||
udi_gap = next((g for g in self.gaps if "UDI System" in g.requirement), None)
|
||||
if udi_gap and udi_gap.status != GapStatus.COMPLETE:
|
||||
recommendations.append(
|
||||
"Implement UDI system and plan for EUDAMED registration. "
|
||||
"Required for placing device on EU market."
|
||||
)
|
||||
|
||||
return recommendations
|
||||
|
||||
|
||||
def format_text_output(result: GapAnalysisResult) -> str:
|
||||
"""Format analysis result as text."""
|
||||
lines = [
|
||||
"=" * 60,
|
||||
"MDR 2017/745 GAP ANALYSIS REPORT",
|
||||
"=" * 60,
|
||||
f"Device: {result.device_name}",
|
||||
f"Class: {result.device_class}",
|
||||
f"Date: {result.analysis_date[:10]}",
|
||||
"",
|
||||
"-" * 60,
|
||||
"SUMMARY",
|
||||
"-" * 60,
|
||||
f"Total Requirements: {result.total_requirements}",
|
||||
f"Gaps Identified: {result.gaps_identified}",
|
||||
f"Completion: {result.completion_percentage}%",
|
||||
"",
|
||||
]
|
||||
|
||||
if result.critical_gaps:
|
||||
lines.extend([
|
||||
"-" * 60,
|
||||
"CRITICAL GAPS (Address Immediately)",
|
||||
"-" * 60,
|
||||
])
|
||||
for gap in result.critical_gaps:
|
||||
lines.append(f" * {gap}")
|
||||
lines.append("")
|
||||
|
||||
lines.extend([
|
||||
"-" * 60,
|
||||
"GAP DETAILS BY CATEGORY",
|
||||
"-" * 60,
|
||||
])
|
||||
|
||||
# Group by category
|
||||
categories = {}
|
||||
for gap in result.gaps:
|
||||
cat = gap["category"]
|
||||
if cat not in categories:
|
||||
categories[cat] = []
|
||||
categories[cat].append(gap)
|
||||
|
||||
for category, gaps in categories.items():
|
||||
lines.append(f"\n{category}:")
|
||||
for gap in gaps:
|
||||
status_mark = "✓" if gap["status"] == "Complete" else "○"
|
||||
lines.append(f" [{status_mark}] {gap['requirement']} ({gap['priority']})")
|
||||
|
||||
lines.extend([
|
||||
"",
|
||||
"-" * 60,
|
||||
"RECOMMENDATIONS",
|
||||
"-" * 60,
|
||||
])
|
||||
for i, rec in enumerate(result.recommendations, 1):
|
||||
lines.append(f"{i}. {rec}")
|
||||
|
||||
lines.append("=" * 60)
|
||||
return "\n".join(lines)
|
||||
|
||||
|
||||
def interactive_mode():
|
||||
"""Run interactive gap analysis session."""
|
||||
print("=" * 60)
|
||||
print("MDR 2017/745 Gap Analysis - Interactive Mode")
|
||||
print("=" * 60)
|
||||
|
||||
device_name = input("\nDevice name: ").strip()
|
||||
if not device_name:
|
||||
device_name = "Unnamed Device"
|
||||
|
||||
print("\nDevice classes:")
|
||||
print(" 1. Class I")
|
||||
print(" 2. Class I (sterile)")
|
||||
print(" 3. Class I (measuring)")
|
||||
print(" 4. Class IIa")
|
||||
print(" 5. Class IIb")
|
||||
print(" 6. Class III")
|
||||
|
||||
class_map = {
|
||||
"1": DeviceClass.I,
|
||||
"2": DeviceClass.I_STERILE,
|
||||
"3": DeviceClass.I_MEASURING,
|
||||
"4": DeviceClass.IIA,
|
||||
"5": DeviceClass.IIB,
|
||||
"6": DeviceClass.III,
|
||||
}
|
||||
|
||||
class_choice = input("\nSelect class (1-6): ").strip()
|
||||
device_class = class_map.get(class_choice, DeviceClass.IIA)
|
||||
|
||||
analyzer = MDRGapAnalyzer(device_name, device_class)
|
||||
|
||||
print("\nFor each requirement, enter status:")
|
||||
print(" c = Complete")
|
||||
print(" i = In Progress")
|
||||
print(" n = Not Started (default)")
|
||||
print(" x = Not Applicable")
|
||||
print(" Enter = Skip (Not Started)")
|
||||
print("")
|
||||
|
||||
status_map = {
|
||||
"c": GapStatus.COMPLETE,
|
||||
"i": GapStatus.IN_PROGRESS,
|
||||
"n": GapStatus.NOT_STARTED,
|
||||
"x": GapStatus.NOT_APPLICABLE,
|
||||
}
|
||||
|
||||
for gap in analyzer.gaps:
|
||||
if gap.status == GapStatus.NOT_APPLICABLE:
|
||||
continue
|
||||
status_input = input(f"{gap.requirement} [c/i/n/x]: ").strip().lower()
|
||||
if status_input in status_map:
|
||||
gap.status = status_map[status_input]
|
||||
|
||||
result = analyzer.analyze()
|
||||
print("\n" + format_text_output(result))
|
||||
|
||||
|
||||
def main():
|
||||
parser = argparse.ArgumentParser(
|
||||
description="EU MDR 2017/745 Gap Analysis Tool"
|
||||
)
|
||||
parser.add_argument("--device", type=str, help="Device name")
|
||||
parser.add_argument(
|
||||
"--class",
|
||||
dest="device_class",
|
||||
choices=["I", "Is", "Im", "IIa", "IIb", "III"],
|
||||
help="Device classification"
|
||||
)
|
||||
parser.add_argument(
|
||||
"--output",
|
||||
choices=["text", "json"],
|
||||
default="text",
|
||||
help="Output format"
|
||||
)
|
||||
parser.add_argument(
|
||||
"--interactive",
|
||||
action="store_true",
|
||||
help="Run in interactive mode"
|
||||
)
|
||||
|
||||
args = parser.parse_args()
|
||||
|
||||
if args.interactive:
|
||||
interactive_mode()
|
||||
return
|
||||
|
||||
if not args.device or not args.device_class:
|
||||
parser.print_help()
|
||||
print("\nError: --device and --class required (or use --interactive)")
|
||||
sys.exit(1)
|
||||
|
||||
class_map = {
|
||||
"I": DeviceClass.I,
|
||||
"Is": DeviceClass.I_STERILE,
|
||||
"Im": DeviceClass.I_MEASURING,
|
||||
"IIa": DeviceClass.IIA,
|
||||
"IIb": DeviceClass.IIB,
|
||||
"III": DeviceClass.III,
|
||||
}
|
||||
|
||||
analyzer = MDRGapAnalyzer(args.device, class_map[args.device_class])
|
||||
result = analyzer.analyze()
|
||||
|
||||
if args.output == "json":
|
||||
print(json.dumps(asdict(result), indent=2))
|
||||
else:
|
||||
print(format_text_output(result))
|
||||
|
||||
|
||||
if __name__ == "__main__":
|
||||
main()
|
||||