name: "risk-management-specialist"
|
||
description: Medical device risk management specialist implementing ISO 14971 throughout product lifecycle. Provides risk analysis, risk evaluation, risk control, and post-production information analysis. Use when user mentions risk management, ISO 14971, risk analysis, FMEA, fault tree analysis, hazard identification, risk control, risk matrix, benefit-risk analysis, residual risk, risk acceptability, or post-market risk.
|
||
---
|
||
|
||
# Risk Management Specialist
|
||
|
||
ISO 14971:2019 risk management implementation throughout the medical device lifecycle.
|
||
|
||
---
|
||
|
||
## Table of Contents
|
||
|
||
- [Risk Management Planning Workflow](#risk-management-planning-workflow)
|
||
- [Risk Analysis Workflow](#risk-analysis-workflow)
|
||
- [Risk Evaluation Workflow](#risk-evaluation-workflow)
|
||
- [Risk Control Workflow](#risk-control-workflow)
|
||
- [Post-Production Risk Management](#post-production-risk-management)
|
||
- [Risk Assessment Templates](#risk-assessment-templates)
|
||
- [Decision Frameworks](#decision-frameworks)
|
||
- [Tools and References](#tools-and-references)
|
||
|
||
---
|
||
|
||
## Risk Management Planning Workflow
|
||
|
||
Establish risk management process per ISO 14971.
|
||
|
||
### Workflow: Create Risk Management Plan
|
||
|
||
1. Define scope of risk management activities:
|
||
- Medical device identification
|
||
- Lifecycle stages covered
|
||
- Applicable standards and regulations
|
||
2. Establish risk acceptability criteria:
|
||
- Define probability categories (P1-P5)
|
||
- Define severity categories (S1-S5)
|
||
- Create risk matrix with acceptance thresholds
|
||
3. Assign responsibilities:
|
||
- Risk management lead
|
||
- Subject matter experts
|
||
- Approval authorities
|
||
4. Define verification activities:
|
||
- Methods for control verification
|
||
- Acceptance criteria
|
||
5. Plan production and post-production activities:
|
||
- Information sources
|
||
- Review triggers
|
||
- Update procedures
|
||
6. Obtain plan approval
|
||
7. Establish risk management file
|
||
8. **Validation:** Plan approved; acceptability criteria defined; responsibilities assigned; file established
|
||
|
||
### Risk Management Plan Content
|
||
|
||
| Section | Content | Evidence |
|
||
|---------|---------|----------|
|
||
| Scope | Device and lifecycle coverage | Scope statement |
|
||
| Criteria | Risk acceptability matrix | Risk matrix document |
|
||
| Responsibilities | Roles and authorities | RACI chart |
|
||
| Verification | Methods and acceptance | Verification plan |
|
||
| Production/Post-Production | Monitoring activities | Surveillance plan |
|
||
|
||
### Risk Acceptability Matrix (5x5)
|
||
|
||
| Probability \ Severity | Negligible | Minor | Serious | Critical | Catastrophic |
|
||
|------------------------|------------|-------|---------|----------|--------------|
|
||
| **Frequent (P5)** | Medium | High | High | Unacceptable | Unacceptable |
|
||
| **Probable (P4)** | Medium | Medium | High | High | Unacceptable |
|
||
| **Occasional (P3)** | Low | Medium | Medium | High | High |
|
||
| **Remote (P2)** | Low | Low | Medium | Medium | High |
|
||
| **Improbable (P1)** | Low | Low | Low | Medium | Medium |
|
||
|
||
### Risk Level Actions
|
||
|
||
| Level | Acceptable | Action Required |
|
||
|-------|------------|-----------------|
|
||
| Low | Yes | Document and accept |
|
||
| Medium | ALARP | Reduce if practicable; document rationale |
|
||
| High | ALARP | Reduction required; demonstrate ALARP |
|
||
| Unacceptable | No | Design change mandatory |
|
||
|
||
---
|
||
|
||
## Risk Analysis Workflow
|
||
|
||
Identify hazards and estimate risks systematically.
|
||
|
||
### Workflow: Conduct Risk Analysis
|
||
|
||
1. Define intended use and reasonably foreseeable misuse:
|
||
- Medical indication
|
||
- Patient population
|
||
- User population
|
||
- Use environment
|
||
2. Select analysis method(s):
|
||
- FMEA for component/function analysis
|
||
- FTA for system-level analysis
|
||
- HAZOP for process deviations
|
||
- Use Error Analysis for user interaction
|
||
3. Identify hazards by category:
|
||
- Energy hazards (electrical, mechanical, thermal)
|
||
- Biological hazards (bioburden, biocompatibility)
|
||
- Chemical hazards (residues, leachables)
|
||
- Operational hazards (software, use errors)
|
||
4. Determine hazardous situations:
|
||
- Sequence of events
|
||
- Foreseeable misuse scenarios
|
||
- Single fault conditions
|
||
5. Estimate probability of harm (P1-P5)
|
||
6. Estimate severity of harm (S1-S5)
|
||
7. Document in hazard analysis worksheet
|
||
8. **Validation:** All hazard categories addressed; all hazards documented; probability and severity assigned
|
||
|
||
### Hazard Categories Checklist
|
||
|
||
| Category | Examples | Analyzed |
|
||
|----------|----------|----------|
|
||
| Electrical | Shock, burns, interference | ☐ |
|
||
| Mechanical | Crushing, cutting, entrapment | ☐ |
|
||
| Thermal | Burns, tissue damage | ☐ |
|
||
| Radiation | Ionizing, non-ionizing | ☐ |
|
||
| Biological | Infection, biocompatibility | ☐ |
|
||
| Chemical | Toxicity, irritation | ☐ |
|
||
| Software | Incorrect output, timing | ☐ |
|
||
| Use Error | Misuse, perception, cognition | ☐ |
|
||
| Environment | EMC, mechanical stress | ☐ |
|
||
|
||
### Analysis Method Selection
|
||
|
||
| Situation | Recommended Method |
|
||
|-----------|-------------------|
|
||
| Component failures | FMEA |
|
||
| System-level failure | FTA |
|
||
| Process deviations | HAZOP |
|
||
| User interaction | Use Error Analysis |
|
||
| Software behavior | Software FMEA |
|
||
| Early design phase | PHA |
|
||
|
||
### Probability Criteria
|
||
|
||
| Level | Name | Description | Frequency |
|
||
|-------|------|-------------|-----------|
|
||
| P5 | Frequent | Expected to occur | >10⁻³ |
|
||
| P4 | Probable | Likely to occur | 10⁻³ to 10⁻⁴ |
|
||
| P3 | Occasional | May occur | 10⁻⁴ to 10⁻⁵ |
|
||
| P2 | Remote | Unlikely | 10⁻⁵ to 10⁻⁶ |
|
||
| P1 | Improbable | Very unlikely | <10⁻⁶ |
|
||
|
||
### Severity Criteria
|
||
|
||
| Level | Name | Description | Harm |
|
||
|-------|------|-------------|------|
|
||
| S5 | Catastrophic | Death | Death |
|
||
| S4 | Critical | Permanent impairment | Irreversible injury |
|
||
| S3 | Serious | Injury requiring intervention | Reversible injury |
|
||
| S2 | Minor | Temporary discomfort | No treatment needed |
|
||
| S1 | Negligible | Inconvenience | No injury |
|
||
|
||
See: [references/risk-analysis-methods.md](references/risk-analysis-methods.md)
|
||
|
||
---
|
||
|
||
## Risk Evaluation Workflow
|
||
|
||
Evaluate risks against acceptability criteria.
|
||
|
||
### Workflow: Evaluate Identified Risks
|
||
|
||
1. Calculate initial risk level from probability × severity
|
||
2. Compare to risk acceptability criteria
|
||
3. For each risk, determine:
|
||
- Acceptable: Document and accept
|
||
- ALARP: Proceed to risk control
|
||
- Unacceptable: Mandatory risk control
|
||
4. Document evaluation rationale
|
||
5. Identify risks requiring benefit-risk analysis
|
||
6. Complete benefit-risk analysis if applicable
|
||
7. Compile risk evaluation summary
|
||
8. **Validation:** All risks evaluated; acceptability determined; rationale documented
|
||
|
||
### Risk Evaluation Decision Tree
|
||
|
||
```
|
||
Risk Estimated
|
||
│
|
||
▼
|
||
Apply Acceptability Criteria
|
||
│
|
||
├── Low Risk ──────────► Accept and document
|
||
│
|
||
├── Medium Risk ───────► Consider risk reduction
|
||
│ │ Document ALARP if not reduced
|
||
│ ▼
|
||
│ Practicable to reduce?
|
||
│ │
|
||
│ Yes──► Implement control
|
||
│ No───► Document ALARP rationale
|
||
│
|
||
├── High Risk ─────────► Risk reduction required
|
||
│ │ Must demonstrate ALARP
|
||
│ ▼
|
||
│ Implement control
|
||
│ Verify residual risk
|
||
│
|
||
└── Unacceptable ──────► Design change mandatory
|
||
Cannot proceed without control
|
||
```
|
||
|
||
### ALARP Demonstration Requirements
|
||
|
||
| Criterion | Evidence Required |
|
||
|-----------|-------------------|
|
||
| Technical feasibility | Analysis of alternative controls |
|
||
| Proportionality | Cost-benefit of further reduction |
|
||
| State of the art | Comparison to similar devices |
|
||
| Stakeholder input | Clinical/user perspectives |
|
||
|
||
### Benefit-Risk Analysis Triggers
|
||
|
||
| Situation | Benefit-Risk Required |
|
||
|-----------|----------------------|
|
||
| Residual risk remains high | Yes |
|
||
| No feasible risk reduction | Yes |
|
||
| Novel device | Yes |
|
||
| Unacceptable risk with clinical benefit | Yes |
|
||
| All risks low | No |
|
||
|
||
---
|
||
|
||
## Risk Control Workflow
|
||
|
||
Implement and verify risk control measures.
|
||
|
||
### Workflow: Implement Risk Controls
|
||
|
||
1. Identify risk control options:
|
||
- Inherent safety by design (Priority 1)
|
||
- Protective measures in device (Priority 2)
|
||
- Information for safety (Priority 3)
|
||
2. Select optimal control following hierarchy
|
||
3. Analyze control for new hazards introduced
|
||
4. Document control in design requirements
|
||
5. Implement control in design
|
||
6. Develop verification protocol
|
||
7. Execute verification and document results
|
||
8. Evaluate residual risk with control in place
|
||
9. **Validation:** Control implemented; verification passed; residual risk acceptable; no unaddressed new hazards
|
||
|
||
### Risk Control Hierarchy
|
||
|
||
| Priority | Control Type | Examples | Effectiveness |
|
||
|----------|--------------|----------|---------------|
|
||
| 1 | Inherent Safety | Eliminate hazard, fail-safe design | Highest |
|
||
| 2 | Protective Measures | Guards, alarms, automatic shutdown | High |
|
||
| 3 | Information | Warnings, training, IFU | Lower |
|
||
|
||
### Risk Control Option Analysis Template
|
||
|
||
```
|
||
RISK CONTROL OPTION ANALYSIS
|
||
|
||
Hazard ID: H-[XXX]
|
||
Hazard: [Description]
|
||
Initial Risk: P[X] × S[X] = [Level]
|
||
|
||
OPTIONS CONSIDERED:
|
||
| Option | Control Type | New Hazards | Feasibility | Selected |
|
||
|--------|--------------|-------------|-------------|----------|
|
||
| 1 | [Type] | [Yes/No] | [H/M/L] | [Yes/No] |
|
||
| 2 | [Type] | [Yes/No] | [H/M/L] | [Yes/No] |
|
||
|
||
SELECTED CONTROL: Option [X]
|
||
Rationale: [Justification for selection]
|
||
|
||
IMPLEMENTATION:
|
||
- Requirement: [REQ-XXX]
|
||
- Design Document: [Reference]
|
||
|
||
VERIFICATION:
|
||
- Method: [Test/Analysis/Review]
|
||
- Protocol: [Reference]
|
||
- Acceptance Criteria: [Criteria]
|
||
```
|
||
|
||
### Risk Control Verification Methods
|
||
|
||
| Method | When to Use | Evidence |
|
||
|--------|-------------|----------|
|
||
| Test | Quantifiable performance | Test report |
|
||
| Inspection | Physical presence | Inspection record |
|
||
| Analysis | Design calculation | Analysis report |
|
||
| Review | Documentation check | Review record |
|
||
|
||
### Residual Risk Evaluation
|
||
|
||
| After Control | Action |
|
||
|---------------|--------|
|
||
| Acceptable | Document, proceed |
|
||
| ALARP achieved | Document rationale, proceed |
|
||
| Still unacceptable | Additional control or design change |
|
||
| New hazard introduced | Analyze and control new hazard |
|
||
|
||
---
|
||
|
||
## Post-Production Risk Management
|
||
|
||
Monitor and update risk management throughout product lifecycle.
|
||
|
||
### Workflow: Post-Production Risk Monitoring
|
||
|
||
1. Identify information sources:
|
||
- Customer complaints
|
||
- Service reports
|
||
- Vigilance/adverse events
|
||
- Literature monitoring
|
||
- Clinical studies
|
||
2. Establish collection procedures
|
||
3. Define review triggers:
|
||
- New hazard identified
|
||
- Increased frequency of known hazard
|
||
- Serious incident
|
||
- Regulatory feedback
|
||
4. Analyze incoming information for risk relevance
|
||
5. Update risk management file as needed
|
||
6. Communicate significant findings
|
||
7. Conduct periodic risk management review
|
||
8. **Validation:** Information sources monitored; file current; reviews completed per schedule
|
||
|
||
### Information Sources
|
||
|
||
| Source | Information Type | Review Frequency |
|
||
|--------|------------------|------------------|
|
||
| Complaints | Use issues, failures | Continuous |
|
||
| Service | Field failures, repairs | Monthly |
|
||
| Vigilance | Serious incidents | Immediate |
|
||
| Literature | Similar device issues | Quarterly |
|
||
| Regulatory | Authority feedback | As received |
|
||
| Clinical | PMCF data | Per plan |
|
||
|
||
### Risk Management File Update Triggers
|
||
|
||
| Trigger | Response Time | Action |
|
||
|---------|---------------|--------|
|
||
| Serious incident | Immediate | Full risk review |
|
||
| New hazard identified | 30 days | Risk analysis update |
|
||
| Trend increase | 60 days | Trend analysis |
|
||
| Design change | Before implementation | Impact assessment |
|
||
| Standards update | Per transition period | Gap analysis |
|
||
|
||
### Periodic Review Requirements
|
||
|
||
| Review Element | Frequency |
|
||
|----------------|-----------|
|
||
| Risk management file completeness | Annual |
|
||
| Risk control effectiveness | Annual |
|
||
| Post-market information analysis | Quarterly |
|
||
| Risk-benefit conclusions | Annual or on new data |
|
||
|
||
---
|
||
|
||
## Risk Assessment Templates
|
||
→ See references/risk-assessment-templates.md for details
|
||
|
||
## Decision Frameworks
|
||
|
||
### Risk Control Selection
|
||
|
||
```
|
||
What is the risk level?
|
||
│
|
||
├── Unacceptable ──► Can hazard be eliminated?
|
||
│ │
|
||
│ Yes─┴─No
|
||
│ │ │
|
||
│ ▼ ▼
|
||
│ Eliminate Can protective
|
||
│ hazard measure reduce?
|
||
│ │
|
||
│ Yes─┴─No
|
||
│ │ │
|
||
│ ▼ ▼
|
||
│ Add Add warning
|
||
│ protection + training
|
||
│
|
||
└── High/Medium ──► Apply hierarchy
|
||
starting at Level 1
|
||
```
|
||
|
||
### New Hazard Analysis
|
||
|
||
| Question | If Yes | If No |
|
||
|----------|--------|-------|
|
||
| Does control introduce new hazard? | Analyze new hazard | Proceed |
|
||
| Is new risk higher than original? | Reject control option | Acceptable trade-off |
|
||
| Can new hazard be controlled? | Add control | Reject control option |
|
||
|
||
### Risk Acceptability Decision
|
||
|
||
| Condition | Decision |
|
||
|-----------|----------|
|
||
| All risks Low | Acceptable |
|
||
| Medium risks with ALARP | Acceptable |
|
||
| High risks with ALARP documented | Acceptable if benefits outweigh |
|
||
| Any Unacceptable residual | Not acceptable - redesign |
|
||
|
||
---
|
||
|
||
## Tools and References
|
||
|
||
### Scripts
|
||
|
||
| Tool | Purpose | Usage |
|
||
|------|---------|-------|
|
||
| [risk_matrix_calculator.py](scripts/risk_matrix_calculator.py) | Calculate risk levels and FMEA RPN | `python risk_matrix_calculator.py --help` |
|
||
|
||
**Risk Matrix Calculator Features:**
|
||
- ISO 14971 5x5 risk matrix calculation
|
||
- FMEA RPN (Risk Priority Number) calculation
|
||
- Interactive mode for guided assessment
|
||
- Display risk criteria definitions
|
||
- JSON output for integration
|
||
|
||
### References
|
||
|
||
| Document | Content |
|
||
|----------|---------|
|
||
| [iso14971-implementation-guide.md](references/iso14971-implementation-guide.md) | Complete ISO 14971:2019 implementation with templates |
|
||
| [risk-analysis-methods.md](references/risk-analysis-methods.md) | FMEA, FTA, HAZOP, Use Error Analysis methods |
|
||
|
||
### Quick Reference: ISO 14971 Process
|
||
|
||
| Stage | Key Activities | Output |
|
||
|-------|----------------|--------|
|
||
| Planning | Define scope, criteria, responsibilities | Risk Management Plan |
|
||
| Analysis | Identify hazards, estimate risk | Hazard Analysis |
|
||
| Evaluation | Compare to criteria, ALARP assessment | Risk Evaluation |
|
||
| Control | Implement hierarchy, verify | Risk Control Records |
|
||
| Residual | Overall assessment, benefit-risk | Risk Management Report |
|
||
| Production | Monitor, review, update | Updated RM File |
|
||
|
||
---
|
||
|
||
## Related Skills
|
||
|
||
| Skill | Integration Point |
|
||
|-------|-------------------|
|
||
| [quality-manager-qms-iso13485](../quality-manager-qms-iso13485/) | QMS integration |
|
||
| [capa-officer](../capa-officer/) | Risk-based CAPA |
|
||
| [regulatory-affairs-head](../regulatory-affairs-head/) | Regulatory submissions |
|
||
| [quality-documentation-manager](../quality-documentation-manager/) | Risk file management |
|